SU-E-T-245: Detection of the Photon Target Damage in Varian Linac Based On Periodical Quality Assurance Measurements
Abstract
Purpose: To determine the best dosimetric metric measured by our routine QA devices for diagnosing photon target failure on a Varian C‐series linac. Methods: We have retrospectively reviewed and analyzed the dosimetry data from a Varian linac with a target degradation that was undiagnosed for one year. A failure in the daily QA symmetry tests was the first indication of an issue. The beam was steered back to a symmetric shape and water scans...
Paper Details
Title
SU-E-T-245: Detection of the Photon Target Damage in Varian Linac Based On Periodical Quality Assurance Measurements
Published Date
Jun 1, 2015
Journal
Volume
42
Issue
6Part15
Pages
3389 - 3389
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History