THREE-DIMENSIONAL DOSIMETRY OF MAMMOGRAPHY USING A BLOCK CUBE BREAST PHANTOM AND RADIOCHROMIC FILM

Volume: 196, Issue: 3-4, Pages: 226 - 233
Published: Oct 20, 2021
Abstract
This study developed a phantom with a shape similar to that of the breast and use GAFCHROMIC films that can be placed inside the phantom to measure the detailed breast dose distribution in mammography. GAFCHROMIC EBT3 was placed on the block cube breast phantom and irradiated with a mammography device to measure the absorbed dose distribution inside the phantom in the horizontal and depth directions. The dose distribution in the horizontal plane...
Paper Details
Title
THREE-DIMENSIONAL DOSIMETRY OF MAMMOGRAPHY USING A BLOCK CUBE BREAST PHANTOM AND RADIOCHROMIC FILM
Published Date
Oct 20, 2021
Volume
196
Issue
3-4
Pages
226 - 233