Gel and thermoluminescence dosimetry for dose verifications of a real anatomy simulated prostate conformal radiation treatment in the presence of metallic femoral prosthesis

Volume: 22, Issue: 10, Pages: 278 - 287
Published: Aug 26, 2021
Abstract
This study aims to verify the dose delivery of prostate radiotherapy treatments in an adult pelvic phantom with two metallic hip and femur prosthesis using a four-field box technique. The prostate planned target volume (PTV) tridimensional (3D) dose distribution was evaluated using gel dosimetry, and thermoluminescent dosimeters (TLD) were used for point-dose measurements outside it. Both results were compared to the treatment planning system...