Technical evaluation of a prototype ratio 29:1 grid for adult patient cardiovascular angiography imaging conditions

Volume: 66, Issue: 14, Pages: 145010 - 145010
Published: Jul 13, 2021
Abstract
The purpose of this work was to assess technical performance of a prototype high-ratio (r29), 80 line cm-1grid for imaging conditions which mimic those for adult cardiovascular angiography. The standard equipment r15, 80 line cm-1grid was used as a reference. Plastic Water®LR phantoms with thickness in the range 20-44 cm were used to simulate adult patient attenuation and scatter. Grids were tested using x-ray field of view 20 and 25 cm and...