Measurement of isocenter alignment accuracy and image distortion of an 0.35 T MR-Linac system

Volume: 64, Issue: 20, Pages: 205011 - 205011
Published: Oct 1, 2019
Abstract
For hybrid devices combining magnetic resonance (MR) imaging and a linac for radiation treatment, the isocenter accuracy as well as image distortions have to be checked. This study presents a new phantom to investigate MR-Linacs in a single measurement in terms of (i) isocentricity of the irradiation and (ii) alignment of the irradiation and imaging isocenter relative to each other using polymer dosimetry gel as well as (iii) 3-dimensional (3D)...
Paper Details
Title
Measurement of isocenter alignment accuracy and image distortion of an 0.35 T MR-Linac system
Published Date
Oct 1, 2019
Volume
64
Issue
20
Pages
205011 - 205011