IMRT national audit in Portugal

Published on Sep 1, 2019in Physica Medica2.485
· DOI :10.1016/J.EJMP.2019.08.013
Tania Santos4
Estimated H-index: 4
(UC: University of Coimbra),
Maria do Carmo Lopes39
Estimated H-index: 39
+ 20 AuthorsJoanna Izewska18
Estimated H-index: 18
(IAEA: International Atomic Energy Agency)
Sources
Abstract
Abstract Purpose The IAEA newly developed “end-to-end” audit methodology for on-site verification of IMRT dose delivery has been carried out in Portugal in 2018. The main goal was to evaluate the physical aspects of the head and neck (H&N) cancer IMRT treatments. This paper presents the national results. Methods All institutions performing IMRT treatments in Portugal, 20 out of 24, have voluntarily participated in this audit. Following the adopted methodology, a Shoulder, Head and Neck End-to-End phantom (SHANE) – that mimics an H&N region, underwent all steps of an IMRT treatment, according to the local practices. The measurements using an ionization chamber placed inside the SHANE phantom at four reference locations (three in PTVs and one in the spinal cord) and an EBT3 film positioned in a coronal plane were compared with calculated doses. FilmQA Pro software was used for film analysis. Results For ionization chamber measurements, the percent difference was within the specified tolerances of ±5% for PTVs and ±7% for the spinal cord in all participating institutions. Considering film analysis, gamma passing rates were on average 96.9%±2.9% for a criterion of 3%/3 mm, 20% threshold, all above the acceptance limit of 90%. Conclusions The national results of the H&N IMRT audit showed a compliance between the planned and the delivered doses within the specified tolerances, confirming no major reasons for concern. At the same time the audit identified factors that contributed to increased uncertainties in the IMRT dose delivery in some institutions resulting in recommendations for quality improvement.
📖 Papers frequently viewed together
12 Authors (Naoki Hayashi)
13 Authors (Julia Henderson)
References34
Newest
#1A. VeresH-Index: 1
#2J.X. HalletH-Index: 1
#1Pavel KazantsevH-Index: 4
#2Victor HernandezH-Index: 12
Last. Joanna IzewskaH-Index: 18
view all 4 authors...
Source
#1Minoru NakaoH-Index: 5
#2Shuichi Ozawa (Hiroshima University)H-Index: 13
Last. Yasushi Nagata (Hiroshima University)H-Index: 47
view all 16 authors...
The accuracy of computed tomography number to electron density (CT-ED) calibration is a key component for dose calculations in an inhomogeneous medium. In a previous work, it was shown that the tolerance levels of CT-ED calibration became stricter with an increase in tissue thickness and decrease in the effective energy of a photon beam. For the last decade, a low effective energy photon beam (e.g., flattening-filter-free (FFF)) has been used in clinical sites. However, its tolerance level has n...
Source
#1Joanna Izewska (IAEA: International Atomic Energy Agency)H-Index: 18
#2Wolfgang Lechner (IAEA: International Atomic Energy Agency)H-Index: 9
Last. Paulina Wesolowska (IAEA: International Atomic Energy Agency)H-Index: 6
view all 3 authors...
Abstract The International Atomic Energy Agency (IAEA) has established a database describing activities of dosimetry audit networks (DAN) in radiotherapy. Since 2010 the data on different aspects of the dosimetry audit have been collected. This information has allowed for the analysis and comparison of current practices in dosimetry auditing activities worldwide. Overall, 79 organizations in 58 countries confirmed that they offer dosimetry audit services for radiotherapy; however, access of radi...
Source
Source
#1Enrica Seravalli (UU: Utrecht University)H-Index: 14
#2Antonetta C. Houweling (UU: Utrecht University)H-Index: 15
Last. Jeroen B. van de Kamer (NKI-AVL: Netherlands Cancer Institute)H-Index: 17
view all 11 authors...
Abstract Background and purpose Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is detecting possible differences between local QA findings and those of an audit, using the same set of treatment plans. Methods A pre-defined set of clinical plans was devised and imported in t...
Source
#1Wolfgang Lechner (Medical University of Vienna)H-Index: 9
#2Paulina Wesolowska (IAEA: International Atomic Energy Agency)H-Index: 6
Last. Joanna Izewska (IAEA: International Atomic Energy Agency)H-Index: 18
view all 19 authors...
Abstract Background and purpose An audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy was developed and tested through a multinational research group and performed on a national level in five different countries. Materials and methods Centres participating in this study were asked to provide OFs calculated by their TPSs for 10 × 10 cm 2 , 6 × 6 cm 2 , 4 × 4 cm 2 , 3 × 3 cm 2 and 2 × 2 cm 2 field siz...
Source
#1Narges Miri (University of Newcastle)H-Index: 5
#2Joerg Lehmann (University of Newcastle)H-Index: 19
Last. Peter B. Greer (University of Newcastle)H-Index: 36
view all 6 authors...
Abstract Background and Purpose Electronic portal imaging devices (EPIDs) can be used to reconstruct dose inside a virtual phantom. This work aims to study the feasibility of using this method for remote dosimetry auditing of clinical trials. Materials and Methods Six centres participated in an intensity modulated radiotherapy (IMRT) pilot study of this new audit approach. Each centre produced a head and neck (HN) and post-prostatectomy (PP) trial plan and transferred the plans to virtual phanto...
Source
#1Ellen M. Husby (Oslo University Hospital)H-Index: 2
#2Elisabeth Dahl Svendsen (Ahus: Akershus University Hospital)H-Index: 1
Last. Anne Catherine Trægde Martinsen (University of Oslo)H-Index: 1
view all 4 authors...
Quality control (QC) of CT scanners is important to evaluate image quality and radiation dose. Different QC phantoms for testing image quality parameters on CT are commercially available, and Catphan phantoms are widely used for this purpose. More data from measured image quality parameters on CT are necessary to assess test methods, tolerance levels, and test frequencies. The aim of this study was to evaluate the stability of essential image quality parameters for axial and helical scans on one...
Source
#2Victor HernandezH-Index: 12
Last. Nuria JornetH-Index: 16
view all 10 authors...
Abstract Background and purpose We performed a multi-centre intercomparison of VMAT dose planning and pre-treatment verification. The aims were to analyse the dose plans in terms of dosimetric quality and deliverability, and to validate whether in-house pre-treatment verification results agreed with those of an external audit. Materials and methods The nine participating centres encompassed different machines, equipment, and methodologies. Two mock cases (prostate and head and neck) were planned...
Source
Cited By4
Newest
#1Tania SantosH-Index: 4
#2Tiago VenturaH-Index: 7
Last. Maria do Carmo LopesH-Index: 39
view all 3 authors...
Abstract A dosimetry system composed by Gafchromic™ EBT films and a flatbed document scanner is a valuable tool for patient-specific quality assurance (QA) of complex external beam radiotherapy treatment plans. However, it is still not widely used in clinical practice. The need for a comprehensive characterization of the film-scanner performance and the establishment of a complex dosimetry protocol may be one of the causes of this weak adhesion. The purpose of this work is to shed light on how t...
Source
#1Tania Santos (UC: University of Coimbra)H-Index: 4
#2Tiago VenturaH-Index: 7
Last. Maria do Carmo LopesH-Index: 39
view all 5 authors...
PURPOSE: Multiple metrics are proposed to characterize and compare the complexity of helical tomotherapy (HT) plans created for different treatment sites. METHODS: A cohort composed of 208 HT plans from head and neck (105), prostate (51) and brain (52) tumor sites was considered. For each plan, 14 complexity metrics were calculated. Those metrics evaluate the percentage of leaves with small opening times or approaching the projection duration, the percentage of closed leaves, the amount of tongu...
Source
#1Pavel Kazantsev (IAEA: International Atomic Energy Agency)H-Index: 4
#2Wolfgang Lechner (Medical University of Vienna)H-Index: 9
Last. Joanna Izewska (IAEA: International Atomic Energy Agency)H-Index: 18
view all 12 authors...
AbstractBackground: The IAEA has developed and tested an on-site, end-to-end IMRT/VMAT dosimetry audit methodology for head and neck cases using an anthropomorphic phantom. The audit methodology is...
Source
#1Tania Santos (UC: University of Coimbra)H-Index: 4
#2Tiago VenturaH-Index: 7
Last. Maria do Carmo LopesH-Index: 39
view all 3 authors...
Abstract Purpose This work aimed to characterize and compare the complexity of the plans created in the context of a national IMRT/VMAT audit. A plan complexity score is proposed to summarize all the evaluated complexity features. Materials and methods Nine complexity metrics have been computed for the audit plans, evaluating different complexity aspects. An approach based on Principal Component Analysis was followed to explore the correlation between the metrics and derive a smaller set of new ...
Source