285. Dose indices in direct digital mammography: Evaluation on phantom and on a patient sample compared with the structured dose report values

Volume: 56, Pages: 235 - 236
Published: Dec 1, 2018
Abstract
Purpose In DR mammography, comparison of the Entrance Skin Dose (ESD) and the Average Glandular Dose (AGD) data provided by the structured dose report (SR) with the measures performed on phantom and on patient. Methods Two identical mammography systems, Amulet (Fujifilm), working in AEC with W/Rh spectrum and voltage range of 26–32 kV, were taken into account. For a PMMA phantom (18 × 24 cm2, thickness of 20–70 mm) and for a sample of 21...
Paper Details
Title
285. Dose indices in direct digital mammography: Evaluation on phantom and on a patient sample compared with the structured dose report values
Published Date
Dec 1, 2018
Volume
56
Pages
235 - 236
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.