Dose Volume Distribution in Digital Breast Tomosynthesis: A Phantom Study

Volume: 1, Issue: 4, Pages: 322 - 328
Published: Jul 1, 2017
Abstract
Monte Carlo (MC) calculations for breast dosimetry in digital breast tomosynthesis (DBT) require experimental validations. We measured the 3-D dose distribution in breast phantoms, using XR-QA2 radiochromic films, compared to dose maps obtained with a previously validated MC code. Film sheets were positioned at the entrance surface, at the bottom surface as well as at four depths between adjacent slabs in the fiveslabs 50-mm-thick phantoms...
Paper Details
Title
Dose Volume Distribution in Digital Breast Tomosynthesis: A Phantom Study
Published Date
Jul 1, 2017
Volume
1
Issue
4
Pages
322 - 328
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.