Study of feasible and safe condition for total body irradiation using cardiac implantable electronic devices.

Published on Sep 20, 2021in Journal of Radiation Research2.724
· DOI :10.1093/JRR/RRAB088
Hiroaki Matsubara17
Estimated H-index: 17
,
Takatomo Ezura1
Estimated H-index: 1
+ 3 AuthorsMasato Tsuneda7
Estimated H-index: 7
Sources
Abstract
Cardiac implantable electronic devices (CIEDs) were believed to have a tolerance dose and that direct irradiation has to be avoided. Thus, no clinical guidelines have mentioned the feasibility of total body irradiation (TBI) with a CIED directly. The purpose of this work was to study a feasible and safe condition for TBI using a CIED. Eighteen CIEDs were directly irradiated by a 6-MV X-ray beam, where a non-neutron producible beam was employed for the removal of any neutron contribution to CIED malfunction. Irradiation up to 10 Gy in accumulated dose was conducted with a 100-cGy/min dose rate, followed by up to 20 Gy at 200 cGy/min. An irradiation test of whether inappropriate ventricular shock therapy was triggered or not was also performed by using a 6-MV beam of 5, 10, 20 and 40 cGy/min to two CIEDs. No malfunction was observed during irradiation up to 20 Gy at 100 and 200 cGy/min without activation of shock therapy. These results were compared with typical TBI, suggesting that a CIED in TBI will not encounter malfunction because the prescribed dose and the dose rate required for TBI are much safer than those used in this experiment. Several inappropriate shock therapies were, however, observed even at 10 cGy/min if activated. The present result suggested that TBI was feasible and safe if a non-neutron producible beam was employed at low dose-rate without activation of shock therapy, where it was not inconsistent with clinical and non-clinical data in the literature. The feasibility of TBI while using a CIED was discussed for the first time.
📖 Papers frequently viewed together
20126.34Heart Rhythm
20 Authors (Akash Makkar, ..., Frank Pelosi)
References47
Newest
#1Andrea BaehrH-Index: 3
#2Lino ConradsH-Index: 1
Last. Uwe HaverkampH-Index: 17
view all 7 authors...
Abstract Background and objective Purpose of this investigation was to get deeper insight into the impact of different radiation techniques and doses on cardiac implantable electric devices (CIEDs). We aimed to mimic a worst-case scenario with very high doses and external radiation being applied directly on the devices. Methods Radiation was applied on 21 CIEDs as photon or electron therapy with maximum dose of 150 Gy in fractions of 2 -20 Gy. CIEDS were put directly into the beam. Brachytherapy...
Source
#1Kazuhiko Nakamura (Aichi Medical University)H-Index: 2
#2Takahiro Aoyama (Aichi Medical University)H-Index: 3
Last. Masaru Nakamura (Aichi Medical University)H-Index: 4
view all 7 authors...
Cumulative dose effects, which are one of the main causes of errors that occur when an implantable cardiac pacemaker (ICP) is irradiated with ionizing radiation, induce permanent failure in ICPs. Although flattening filter free (FFF) beams, which are often used in stereotactic radiotherapy, are known to have different characteristics from conventional (with flattening filter [WFF]) beams, the cumulative dose effects on ICPs with FFF beams have been under-investigated. This study investigates ICP...
Source
BACKGROUND AND PURPOSE High-dose 6‑MV radiotherapy may cause cardiac implantable electronic devices (CIEDs) to malfunction. To assess CIED malfunctions resulting from direct exposure up to 10 Gy, 100 pacemakers (PMs) and 40 implantable cardioverter-defibrillators (ICDs) were evaluated. MATERIALS AND METHODS CIEDs underwent baseline interrogation. In ICDs, antitachycardia therapies were disabled via the programmer while the detection windows were left enabled. A computed tomography (CT) scan was ...
Source
#1Toshiki Ohno (Dokkyo Medical University)H-Index: 1
#2Toshinori SoejimaH-Index: 21
Last. Eisaku Yoden (Kawasaki Medical School)H-Index: 14
view all 12 authors...
Source
#1Christiana Schernthaner (Paracelsus Private Medical University of Salzburg)H-Index: 8
#2Michael KoppH-Index: 2
Last. Bernhard StrohmerH-Index: 8
view all 9 authors...
Patients with cardiac implantable electronic devices undergoing radiotherapy (RT) are prone to the risk of device failure. Guidelines and manufacturer's instructions are lacking practical recommendations for cumulative radiation doses to pacemakers or implantable cardioverter defibrillators. The present case demonstrates the effect of RT of a Merkel cell carcinoma near the location of a cardiac resynchronization therapy pacemaker. Despite guideline recommendations, surgical relocation or de novo...
Source
#1Szymon MielczarekH-Index: 1
#2Paweł SyskaH-Index: 9
Last. Hanna SzwedH-Index: 12
view all 7 authors...
Source
#1Hossein Aslian (UniTS: University of Trieste)H-Index: 5
#2Tomas Kron (Peter MacCallum Cancer Centre)H-Index: 56
Last. Mara SevergniniH-Index: 8
view all 12 authors...
PURPOSE: This study focused on determining risks from stereotactic radiotherapy using flattening filter-free (FFF) beams for patients with cardiac implantable electronic device (CIEDs). Two strategies were employed: a) a retrospective analysis of patients with CIEDs who underwent stereotactic radiosurgery (SRS)/SBRT at the Peter MacCallum Cancer Centre between 2014 and 2018 and b) an experimental study on the impact of FFF beams on CIEDs. METHODS: A retrospective review was performed. Subsequent...
Source
#1Kazuhiko Nakamura (Aichi Medical University)H-Index: 2
#2Takahiro AoyamaH-Index: 3
Last. Masaru Nakamura (Aichi Medical University)H-Index: 4
view all 7 authors...
Direct irradiation may cause malfunctioning of cardiac implantable electronic devices (CIEDs). Therefore, a treatment plan that does not involve direct irradiation of CIEDs should be formulated. However, CIEDs may be directly exposed to radiation because of the sudden intrafractional movement of the patient. The probability of CIED malfunction reportedly depends on the dose rate; however, reports are only limited to dose rates 8 Gy/min on CIED function. Four CIEDs were placed at the center of th...
Source
#1Hiroaki MatsubaraH-Index: 17
#2Takatomo EzuraH-Index: 1
Last. Masato TsunedaH-Index: 7
view all 6 authors...
PURPOSE: Cardiac implantable electronic devices (CIEDs) were believed to possess a tolerance dose to malfunction during radiotherapy. Although recent studies have qualitatively suggested neutrons as a cause of malfunction, numerical understanding has not been reached. The purpose of this work is to quantitatively clarify the contribution of secondary neutrons from out-of-field irradiation to the malfunction of CIEDs as well as to deduce the frequency of malfunctions until completion of prostate ...
Source
#1Benjamin Gauter-Fleckenstein (Heidelberg University)H-Index: 5
#2Julia Nguyen (Heidelberg University)H-Index: 1
Last. Erol Tülümen (Heidelberg University)H-Index: 12
view all 9 authors...
Abstract Background and purpose Providing evidence for radiotherapy (RT)-induced effects on cardiac implantable electric devices (CIEDs) with focus on flattening filter free-volumetric modulated arc therapy (FFF-VMAT) at 6 and 10 MV as well as 3D-conformal radiotherapy (3D-CRT) at 18 MV. Materials and methods 68 CIEDs (64 implantable cardioverter-defibrillators (ICDs) and 4 cardiac pacemakers (PMs)) were located on the left chest position on a slab phantom and irradiated under telemetrical surve...
Source
Cited By0
Newest
This website uses cookies.
We use cookies to improve your online experience. By continuing to use our website we assume you agree to the placement of these cookies.
To learn more, you can find in our Privacy Policy.