Sensitivity and specificity analysis of 2D small field measurement array: Patient‐specific quality assurance of small target treatments and spatially fractionated radiotherapy

Volume: 22, Issue: 10, Pages: 104 - 119
Published: Aug 27, 2021
Abstract
The aim of this paper is to describe the tests carried out on a SRSMapCheck array, to verify its reliability and sensitivity for quality assurance (QA) of high gradient treatments as an alternative system to the use of high spatial resolution detectors, such as gafchromic film, whose processing requires meticulous and time-consuming procedures.In an initial step, general functionality tests were carried out to verify that the equipment meets the...
Paper Details
Title
Sensitivity and specificity analysis of 2D small field measurement array: Patient‐specific quality assurance of small target treatments and spatially fractionated radiotherapy
Published Date
Aug 27, 2021
Volume
22
Issue
10
Pages
104 - 119
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.