Stress evaluation of different implant lengths on atrophic edentulous mandibles with fixed full-arch implant-supported prosthesis: a finite element analysis

Volume: 24, Issue: 4, Pages: 358 - 374
Published: Mar 12, 2021
Abstract
Finite element analysis was used to compare the effect of different implant lengths on atrophic mandible with full-arch fixed prostheses. Four models were constructed with different implant lengths: 4, 6, 8 and 10 mm. A 100-N occlusal load was applied. The stress at the bone level, implant, and prosthetic components were obtained. Similar behavior was observed for all groups, except for 4 mm, which showed more discrepant values ​​for all...
Paper Details
Title
Stress evaluation of different implant lengths on atrophic edentulous mandibles with fixed full-arch implant-supported prosthesis: a finite element analysis
Published Date
Mar 12, 2021
Volume
24
Issue
4
Pages
358 - 374
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.