Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard

Volume: 7, Issue: 1
Published: Jun 26, 2020
Abstract
Background The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard. Methods According to the NEMA NU2-2018, spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and timing resolution were evaluated. Spatial resolution was measured by using a 22 Na point source. System...
Paper Details
Title
Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard
Published Date
Jun 26, 2020
Volume
7
Issue
1
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.