Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images

Volume: 7, Issue: 1
Published: Mar 17, 2020
Abstract
Evaluate the physical performance of the VERITON CzT camera (Spectrum Dynamics, Caesarea, Israel) that benefits from new detection architecture enabling whole-body imaging compared to that of a conventional dual-head Anger camera.Different line sources and phantom measurements were performed on each system to evaluate spatial resolution, sensitivity, energy resolution and image quality with acquisition and reconstruction parameters similar to...
Paper Details
Title
Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images
Published Date
Mar 17, 2020
Volume
7
Issue
1
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.