Dose reduction for CT coronary calcium scoring with a calcium-aware image reconstruction technique: a phantom study

Published on Feb 19, 2020in European Radiology4.101
· DOI :10.1007/S00330-020-06709-9
Ronald Booij7
Estimated H-index: 7
(EUR: Erasmus University Rotterdam),
Niels R. van der Werf5
Estimated H-index: 5
(UU: Utrecht University)
+ 2 AuthorsMarcel van Straten12
Estimated H-index: 12
(EUR: Erasmus University Rotterdam)
Sources
Abstract
To assess the dose reduction potential of a calcium-aware reconstruction technique, which aims at tube voltage-independent computed tomography (CT) numbers for calcium. A cardiothoracic phantom, mimicking three different patient sizes, was scanned with two calcium inserts (named D100 and CCI), containing calcifications varying in size and density. Tube voltage was varied both manually (range 70–150 and Sn100 kVp) and automatically. Tube current was automatically adapted to maintain reference image quality defined at 120 kVp. Data was reconstructed with the standard reconstruction technique (kernel Qr36) and the calcium-aware reconstruction technique (kernel Sa36). We assessed the radiation dose reduction potential (volumetric CT dose index values (CTDIvol)), noise (standard deviation (SD)), mean CT number (HU) of each calcification, and Agatston scores for varying kVp. Results were compared with the reference acquired at 120 kVp and reconstructed with Qr36. Automatic selection of the optimal tube voltage resulted in a CTDIvol reduction of 22%, 15%, and 12% compared with the reference for the small, medium, and large phantom, respectively. CT numbers differed up to 64% for the standard reconstruction and 11% for the calcium-aware reconstruction. Similarly, Agatston scores deviated up to 40% and 8% for the standard and calcium-aware reconstruction technique, respectively. CT numbers remained consistent with comparable calcium scores when the calcium-aware image reconstruction technique was applied with varying tube voltage. Less consistency was observed in small calcifications with low density. Automatic reduction of tube voltage resulted in a dose reduction of up to 22%. • The calcium-aware image reconstruction technique allows for consistent CT numbers when varying the tube voltage. • Automatic reduction of tube voltage results in a reduced radiation exposure of up to 22%. • This study stresses the known limitations of the current Agatston score technique.
📖 Papers frequently viewed together
32 Citations
10 Citations
37 Citations
References20
Newest
#1Martin J. Willemink (Stanford University)H-Index: 23
#2Niels R. van der Werf (UU: Utrecht University)H-Index: 5
Last. Dominik Fleischmann (Stanford University)H-Index: 55
view all 6 authors...
Abstract Coronary artery calcium (CAC) is a strong predictor for future cardiovascular events. Traditionally CAC has been quantified using the Agatston score, which was developed in the late 1980s for electron beam tomography (EBT). While EBT has been completely replaced by modern multiple-detector row CT technology, the traditional CAC scoring method by Agatston remains in use, although the literature indicates suboptimal reproducibility and subjects being incorrectly classified. The traditiona...
15 CitationsSource
#1Georg Apfaltrer (MUSC: Medical University of South Carolina)H-Index: 3
#2Moritz H. Albrecht (MUSC: Medical University of South Carolina)H-Index: 24
Last. Christian Tesche (MUSC: Medical University of South Carolina)H-Index: 18
view all 16 authors...
Objectives To investigate diagnostic accuracy and radiation dose of high-pitch CT coronary artery calcium scoring (CACS) with tin filtration (Sn100kVp) versus standard 120kVp high-pitch acquisition.
15 CitationsSource
#1Michael J. Blaha (Johns Hopkins University)H-Index: 83
#2Martin Bødtker Mortensen (Aarhus University Hospital)H-Index: 20
Last. Miguel Cainzos-Achirica (Johns Hopkins University)H-Index: 17
view all 5 authors...
Abstract Quantification of coronary artery calcium (CAC) has been shown to be reliable, reproducible, and predictive of cardiovascular risk. Formal CAC scoring was introduced in 1990, with early scoring algorithms notable for their simplicity and elegance. Yet, with little evidence available on how to best build a score, and without a conceptual model guiding score development, these scores were, to a large degree, arbitrary. In this review, we describe the traditional approaches for clinical CA...
103 CitationsSource
#1Christian Tesche (MUSC: Medical University of South Carolina)H-Index: 18
#2Carlo N. De Cecco (MUSC: Medical University of South Carolina)H-Index: 36
Last. John W. Nance (MUSC: Medical University of South Carolina)H-Index: 24
view all 14 authors...
Abstract Objectives To investigate the diagnostic accuracy of CT coronary artery calcium scoring (CACS) with tin pre-filtration (Sn100 kVp) using iterative beam-hardening correction (IBHC) calcium material reconstruction compared to the standard 120 kVp acquisition. Background Third generation dual-source CT (DSCT) CACS with Sn100 kVp acquisition allows significant dose reduction. However, the Sn100 kVp spectrum is harder with lower contrast compared to 120kVp, resulting in lower calcium score v...
12 CitationsSource
#1Marleen Vonder (UMCG: University Medical Center Groningen)H-Index: 7
#2Gert Jan Pelgrim (UMCG: University Medical Center Groningen)H-Index: 9
Last. Rozemarijn Vliegenthart (UMCG: University Medical Center Groningen)H-Index: 24
view all 9 authors...
Objectives To evaluate detectability and quantification of coronary calcifications for CT with a tin filter for spectral shaping.
11 CitationsSource
#1Harvey S. Hecht (ISMMS: Icahn School of Medicine at Mount Sinai)H-Index: 32
#2Michael J. Blaha (Johns Hopkins University)H-Index: 83
Last. Leslee J. Shaw (Emory University)H-Index: 136
view all 10 authors...
Abstract This expert consensus statement summarizes the available data regarding the prognostic value of CAC in the asymptomatic population and its ability to refine individual risk prediction, addresses the limitations identified in the current traditional risk factor-based treatment strategies recommended by the 2013 ACC/AHA Prevention guidelines including use of the Pooled Cohort Equations (PCE), and the US Preventive Services Task Force (USPSTF) Recommendation Statement for Statin Use for th...
151 CitationsSource
#1Harvey S. Hecht (NYMC: New York Medical College)H-Index: 32
#2Paul Cronin (UM: University of Michigan)H-Index: 24
Last. Suhny Abbara (UTSW: University of Texas Southwestern Medical Center)H-Index: 76
view all 8 authors...
The Society of Cardiovascular Computed Tomography (SCCT) and the Society of Thoracic Radiology (STR) have jointly produced this document. Experts in this subject have been selected from both organizations to examine subject-specific data and write this guideline in partnership. A formal literature review, weighing the strength of evidence has been performed. When available, information from studies on cost was considered. Computed tomography (CT) acquisition, CAC scoring methodologies and clinic...
139 CitationsSource
#1Martin J. Willemink (UU: Utrecht University)H-Index: 23
#2Bronislaw Abramiuc (UMCG: University Medical Center Groningen)H-Index: 1
Last. Tim Leiner (UU: Utrecht University)H-Index: 56
view all 11 authors...
Abstract Objective To evaluate the effect of chest size on coronary calcium score (CCS) as assessed with new-generation CT systems from 4 major vendors. Methods An anthropomorphic, small-sized (300 × 200 mm) chest phantom containing 100 small calcifications (diameters, 0.5–2.0 mm) was evaluated with and without an extension ring on state-of-the-art CT systems from 4 vendors. The extension ring was used to mimic a patient with a large chest size (400 × 300 mm). Image acquisition was repeated 5 ti...
11 CitationsSource
#1Koen Nieman (EUR: Erasmus University Rotterdam)H-Index: 64
Advanced coronary atherosclerosis is characterized by calcium deposition. Although coronary calcium had been visualized by fluoroscopy and conventional computed tomography (CT), it was not until the introduction of electron beam computed tomography (EBCT) that high-resolution, motion-free images of the coronary arteries could be acquired. In their seminal paper from 1990, Agatston et al. 1 describe a methodology to image and quantify calcium in the coronary arteries. Cohort studies demonstrated ...
7 CitationsSource
#1Ramit Lamba (UC Davis: University of California, Davis)H-Index: 16
#2John P. McGahanH-Index: 63
Last. John M. BooneH-Index: 74
view all 7 authors...
OBJECTIVE. The purpose of this study is to determine whether Hounsfield numbers of soft tissues on unenhanced abdominal CT of the same patient vary on repeat scans done on two different manufacturers’ MDCT scanners. MATERIALS AND METHODS. A database search was performed to identify patients older than 18 years who underwent unenhanced CT of the abdomen and pelvis performed both on a Volume CT (GE Healthcare) and a Definition AS Plus (Siemens Healthcare) 64-MDCT scanner within 12 months of each o...
66 CitationsSource
Cited By6
Newest
#1Niels R. van der Werf (UU: Utrecht University)H-Index: 5
#2Salim Si-Mohamed (UCBL: Claude Bernard University Lyon 1)H-Index: 13
Last. Martin J. Willemink (Stanford University)H-Index: 23
view all 11 authors...
Objective null The aim of the current study was, first, to assess the coronary artery calcium (CAC) scoring potential of spectral photon-counting CT (SPCCT) in comparison with computed tomography (CT) for routine clinical protocols. Second, improved CAC detection and quantification at reduced slice thickness were assessed. null Methods null Raw data was acquired and reconstructed with several combinations of reduced slice thickness and increasing strengths of iterative reconstruction (IR) for bo...
Source
#1Gijs D. van Praagh (Stanford University)
#2Niels R. van der Werf (UU: Utrecht University)H-Index: 5
Last. Martin J. Willemink (Stanford University)H-Index: 23
view all 10 authors...
OBJECTIVE Coronary artery calcium (CAC) score is a strong predictor for future adverse cardiovascular events. Anthropomorphic phantoms are often used for CAC studies on computed tomography (CT) to allow for evaluation or variation of scanning or reconstruction parameters within or across scanners against a reference standard. This often results in large number of datasets. Manual assessment of these large datasets is time consuming and cumbersome. Therefore, this study aimed to develop and valid...
Source
#1Di Zhang (Cedars-Sinai Medical Center)
#2A Scott (Cedars-Sinai Medical Center)H-Index: 4
Last. Yifang Zhou (Cedars-Sinai Medical Center)H-Index: 11
view all 6 authors...
Abstract Purpose Population dose has been a concern with coronary artery calcium scoring CT since it is performed in adults with borderline risk. Lower tube voltage acquisitions are appealing but there are no agreed schemes for reduced dose determination. Moreover, conventional scoring cannot be used without changing the multiple Agatston thresholds. Methods By applying consistent calcium contrast-to-noise ratio to two anthropomorphic heart phantoms (medium and large) with 3-cm hydroxyapatite (H...
Source
#1Luke G. Terlouw (EUR: Erasmus University Rotterdam)H-Index: 2
Last. Adriaan Moelker (EUR: Erasmus University Rotterdam)H-Index: 27
view all 5 authors...
A practical screening tool for chronic mesenteric ischemia (CMI) could facilitate early recognition and reduce undertreatment and diagnostic delay. This study explored the ability to discriminate CMI from non-CMI patients with a mesenteric artery calcium score (MACS). This retrospective study included CTAs of consecutive patients with suspected CMI in a tertiary referral center between April 2016 and October 2019. A custom-built software module, using the Agatston definition, was developed and u...
1 CitationsSource
#1Niels R. van der Werf (UU: Utrecht University)H-Index: 5
#2Ronald Booij (Erasmus University Medical Center)
Last. Marcel J W Greuter (UT: University of Twente)H-Index: 21
view all 10 authors...
OBJECTIVES The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstruction technique on moving CAC was evaluated with a dynamic phantom. METHODS An artificial coronary artery, containing two CAC of equal size and different densities (196 ± 3, 380 ± 2 mg hydroxyapatite cm-3)...
Source
#1Shengzhen Tao (Mayo Clinic)H-Index: 13
#2Emily N. Sheedy (Mayo Clinic)H-Index: 1
Last. Shuai Leng (Mayo Clinic)H-Index: 53
view all 10 authors...
PURPOSE To determine the accuracy of CT number and calcium score of a kV-independent technique based on an artificial 120 kV reconstruction, and its potential to reduce radiation dose. METHODS Anthropomorphic chest phantoms were scanned on a 3rd-generation dual-source CT system equipped with the artificial 120 kV reconstruction. First, a phantom module containing a 20-mm diameter hydroxyapatite (HA) insert was scanned inside the chest phantoms at different tube potentials (70-140 kV) to evaluate...
1 CitationsSource