Practical implementation of robust MR-thermometry during clinical MR-guided microwave ablations in the liver at 1.5 T
Abstract
Practical non-invasive equipment modifications and effective acquisition methods to achieve robust and reliable real-time MR thermometry for monitoring of clinical hepatic microwave ablations were implemented. These included selection of the microwave generator location (inside versus outside the MR scan room), the number of radiofrequency chokes added to the microwave generator's coaxial lines, and the use of copper wool to maximize their...
Paper Details
Title
Practical implementation of robust MR-thermometry during clinical MR-guided microwave ablations in the liver at 1.5 T
Published Date
Nov 1, 2019
Journal
Volume
67
Pages
91 - 99