Intrasession and Between-Visit Variability of Retinal Vessel Density Values Measured with OCT Angiography in Diabetic Patients.

Published on Jul 13, 2018in Scientific Reports3.998
· DOI :10.1038/S41598-018-28994-7
Cecília Czakó6
Estimated H-index: 6
(Semmelweis University),
Gábor László Sándor12
Estimated H-index: 12
(Semmelweis University)
+ 5 AuthorsIllés Kovács23
Estimated H-index: 23
(Semmelweis University)
Sources
Abstract
In clinical practice the measurement error of an instrument has special importance in analyzing and interpreting data, and acknowledging limitations. The purpose of this study was to evaluate intrasession and between-visit reproducibility of OCT angiography measurements in diabetic patients. A total of 54 eyes of 27 diabetic patients underwent OCT angiography imaging. Foveal avascular zone (FAZ) area and superficial retinal vessel density (VD) at 3 mm were calculated using the AngioAnalytics software. Three consecutive images were acquired at first visit and one image 1 month later. Intrasession and between-visit reproducibility of parameters were characterized by intraclass correlation coefficient (ICC), coefficient of variation (CV), and coefficient of repeatability (CR) values. We measured excellent (>0.90) ICC values both in intrasession and between-visit comparisons. CV was higher for the FAZ area compared to VD both in intrasession (7.79% vs. 2.87%) and in between-visit (12.33% vs. 2.95%) comparisons. Between-visit CR value for VD was 4.53% (95% CI: 3.72–5.79%). These data suggest that OCT angiography shows excellent repeatability in diabetic patients, indicating that this non-invasive technology might be suitable for longitudinal assessment of microvascular complications.
📖 Papers frequently viewed together
10 Citations
100 Citations
481 Citations
References44
Newest
#1Yuanjun Zhao (Penn State Milton S. Hershey Medical Center)H-Index: 5
#2Sarah R. Weber (UM: University of Michigan)H-Index: 6
Last. Jeffrey M. Sundstrom (UM: University of Michigan)H-Index: 14
view all 11 authors...
Purpose: To investigate the molecular components of the vitreous in order to better understand retinal physiology and disease. Methods: Vitreous was acquired from patients undergoing vitrectomy for macular hole and/or epiretinal membrane, postmortem donors, and C57BL/6J mice. Unbiased proteomic analysis was performed via electrospray ionization tandem mass spectrometry (MS/MS). Gene ontology analysis was performed and results were confirmed with transmission electron microscopy, atomic force mic...
35 CitationsSource
#1Beau J. FennerH-Index: 4
#2Gavin TanH-Index: 30
Last. Gemmy CheungH-Index: 22
view all 6 authors...
Purpose Optical coherence tomography angiography (OCT-A) potentially allows for rapid and non-invasive quantification of retinal capillary plexus density in various disease states. This study aims to identify the key features that influenced the repeatability of OCT-A measurements. Methods We obtained OCT-A images on two separate visits in 44 healthy eyes from 44 subjects, each imaged with using the Topcon DRI OCT Triton imaging system. The parafoveal vessel density within a 1.5 mm radius centre...
38 CitationsSource
#2Harsha L. RaoH-Index: 29
Last. Carroll A.B. Webers (UM: Maastricht University)H-Index: 34
view all 11 authors...
Aims To compare the intrasession repeatability of peripapillary and macular vessel density measurements of optical coherence tomography angiography (OCTA) in normal and glaucoma eyes, and to evaluate the effect of signal strength of OCTA scans on the repeatability. Methods In a cross-sectional study, three optic nerve head scans each of 65 eyes (30 normal, 35 glaucoma eyes) and three macular scans each of 69 eyes (35 normal, 34 glaucoma eyes) acquired in the same session with OCTA were analysed....
68 CitationsSource
#1Fred K. Chen (UWA: University of Western Australia)H-Index: 29
#2Moreno Menghini (UWA: University of Western Australia)H-Index: 10
Last. Danuta M. Sampson (UWA: University of Western Australia)H-Index: 4
view all 6 authors...
Purpose: To measure intrasession repeatability and interocular symmetry of the foveal avascular zone area (FAZA) and superficial retinal vessel density (SRVD) using AngioVue Analytics optical coherence tomography angiography (OCTA). Methods: Fifty healthy individuals were prospectively enrolled. OCTA scans (3 × 3 and 6 × 6 mm) were acquired twice in right and once in left eyes. FAZA (with and without rescaling) and SRVD for 18 regions (whole, fovea, parafovea, six parafoveal subregions, and nine...
20 CitationsSource
#1Jianqin Lei (Xi'an Jiaotong University)H-Index: 13
#2Mary K Durbin (ZEISS: Carl Zeiss AG)H-Index: 37
Last. Srinivas R. Sadda (UCLA: University of California, Los Angeles)H-Index: 79
view all 8 authors...
Importance The repeatability and reproducibility of quantitative metrics from optical coherence tomographic angiography (OCTA) must be assessed before these data can be confidently interpreted in clinical research and practice. Objective To evaluate the repeatability and reproducibility of OCTA-derived retinal vascular quantitative metrics. Design, Setting and Participants In this cross-sectional study, 21 healthy volunteers (42 eyes) and 22 patients with retinal disease (22 eyes), including 14 ...
100 CitationsSource
#1Hideki Shiihara (Kadai: Kagoshima University)H-Index: 6
#2Taiji Sakamoto (Kadai: Kagoshima University)H-Index: 60
Last. Shozo Sonoda (Kadai: Kagoshima University)H-Index: 22
view all 7 authors...
This study was performed to compare the area of the foveal avascular zone (FAZ-area) obtained by three optical coherence tomography angiography (OCTA) instruments. This was a cross-sectional, non-interventional study of twenty-seven healthy right eyes. The superficial and deep FAZ-area was measured manually with three OCTA instruments: Triton (Topcon), RS3000 (Nidek), and CIRRUS (Zeiss). The intra-rater, inter-rater, and inter-instrument correlation coefficients (CC) were assessed. The intra-rat...
32 CitationsSource
#1Özge Yanik Odabaş (Ankara University)H-Index: 1
#2Sibel DemirelH-Index: 11
Last. Figen BatıoğluH-Index: 17
view all 4 authors...
Purpose:To evaluate the repeatability of vessel density and superficial and deep foveal avascular zone measurements using optical coherence tomography angiography, and to specify a diurnal change range.Methods:Forty-six eyes of 25 healthy individuals were included. Optical coherence tomography angio
26 CitationsSource
#1Mary K Durbin (ZEISS: Carl Zeiss AG)H-Index: 37
#2Lin An (ZEISS: Carl Zeiss AG)H-Index: 29
Last. José Cunha-VazH-Index: 65
view all 8 authors...
Importance Quantitative measurements based on optical coherence tomographic angiography (OCTA) may have value in managing diabetic retinopathy (DR), but there is limited information on the ability of OCTA to distinguish eyes with DR. Objective To evaluate the ability of measurements of retinal microvasculature using OCTA to distinguish healthy eyes from eyes with DR. Design, Setting, and Participants In this prospective cross-sectional study, OCTA was used to examine the eyes of participants wit...
164 CitationsSource
#1Mayss Al-Sheikh (UCLA: University of California, Los Angeles)H-Index: 17
#2Tudor TepelusH-Index: 12
Last. Srinivas R. Sadda (UCLA: University of California, Los Angeles)H-Index: 79
view all 4 authors...
Aims To determine the repeatability of optical coherence tomography angiography (OCT-A)-derived automated vessel density measurements in the superficial retinal layer (SRL) and deep retinal layer (DRL) in healthy individuals. Methods 41 eyes of 21 healthy individuals were included in this study. OCT-A was performed using the NIDEK RS-3000 Advance with prototype OCT-A software. Each subject underwent two scanning sessions with 5–10 min in between. En face images of the retinal vasculature were ge...
94 CitationsSource
#1Alessio Montuoro (Medical University of Vienna)H-Index: 14
#2Sebastian M Waldstein (Medical University of Vienna)H-Index: 10
Last. Hrvoje Bogunovic (Medical University of Vienna)H-Index: 24
view all 5 authors...
Modern optical coherence tomography (OCT) devices used in ophthalmology acquire steadily increasing amounts of imaging data. Thus, reliable automated quantitative analysis of OCT images is considered to be of utmost importance. Current automated retinal OCT layer segmentation methods work reliably on healthy or mildly diseased retinas, but struggle with the complex interaction of the layers with fluid accumulations in macular edema. In this work, we present a fully automated 3D method which is a...
47 CitationsSource
Cited By16
Newest
Purpose Two-dimensional (2D) optical coherence tomography angiography (OCTA) is known to be prone to segmentation errors, especially in pathologic eyes. Therefore, our aim was to systematically compare intra-session repeatability between repeated scans for 2D and three-dimensional (3D) OCTA metrics in quantifying retinal perfusion in eyes with diabetic macular edema (DME). Methods Diabetic patients with diabetic retinopathy (DR) and DME who had two consecutive OCTA imaging scans obtained during ...
Source
Purpose To analyze the repeatability of vessel density (VD) measurements and manual foveal avascular zone (FAZ) measurements using optical coherence tomography angiography (OCTA) in patients with retinal vein occlusion (RVO) without macular edema. Methods The study population consisted of patients with RVO and central macular thickness l300 μm. For each subject, measurements were performed twice with a 5-minute interval. The intraclass correlation coefficient (ICC) and coefficient of variation (...
Source
#1Lekha Mukkamala (UC Davis: University of California, Davis)H-Index: 1
Last. Susanna S. ParkH-Index: 6
view all 4 authors...
Purpose Although commercial optical coherence tomography angiography (OCTA) machines quantitate retinal vascular density (VD) by dividing the vasculature into superficial and deep capillary plexus (SCP, DCP), histology reveals three distinct plexus layers. This study tested the hypothesis that the VD measurement of three distinct retinal plexus layers obtained using custom segmentation has high repeatability comparable to that of automatically segmented SCP and DCP layers. Materials and methods ...
Source
ABSTRACT The introduction of optical coherence tomography angiography (OCTA) has significantly improved our knowledge on the ocular vascular alterations occurring in diabetes. In this article, we will provide a review of the prominent OCTA findings in diabetes. In details, OCTA demonstrated that both the retinal vessels and the choriocapillaris are affected in diabetic subjects. The recent employment of widefield technology and a 3-dimensional (3D) visualization in OCTA imaging are also discusse...
1 CitationsSource
#1Giuseppe QuerquesH-Index: 54
#2Enrico BorrelliH-Index: 26
Last. Francesco BandelloH-Index: 64
view all 5 authors...
The introduction of optical coherence tomography angiography (OCTA) has remarkably expanded our knowledge of the ocular vascular alterations occurring in diabetes. In this article, a review of the prominent OCTA findings in diabetes is followed by a description of salient histological and anatomical features of microaneurysms, essential for the proper interpretation of in vivo imaging of these retinal vascular abnormalities. The recent employment of a three-dimensional (3D) visualization in OCTA...
1 CitationsSource
#1Cecília Czakó (Semmelweis University)H-Index: 6
#2Tibor Kovács (Semmelweis University)H-Index: 14
Last. Illés Kovács (Semmelweis University)H-Index: 2
view all 14 authors...
Cognitive impairment and dementia are major medical, social, and economic public health issues worldwide with significant implications for life quality in older adults. The leading causes are Alzheimer's disease (AD) and vascular cognitive impairment/dementia (VCID). In both conditions, pathological alterations of the cerebral microcirculation play a critical pathogenic role. Currently, the main pathological biomarkers of AD-β-amyloid peptide and hyperphosphorylated tau proteins-are detected eit...
6 CitationsSource
#1Khalil Ghasemi Falavarjani (IUMS: Iran University of Medical Sciences)H-Index: 30
#2Reza Mirshahi (IUMS: Iran University of Medical Sciences)H-Index: 9
Last. Mahsa Sardarinia (IUMS: Iran University of Medical Sciences)H-Index: 2
view all 4 authors...
Aim:To determine the minimum number of optical coherence tomography B-scan corrections required to provide acceptable vessel density measurements on optical coherence tomography angiography images ...
1 CitationsSource
#1Jimmy Hong (NTU: Nanyang Technological University)H-Index: 3
#2Bingyao TanH-Index: 11
Last. Jacqueline Chua (NUS: National University of Singapore)H-Index: 17
view all 10 authors...
PURPOSE: To assess the repeatability of retinal vascular metrics using different postprocessing methods as obtained from the swept-source optical coherence tomography angiography (SS-OCTA). METHODS: Thirty-two participants (63% males; mean [SD] age, 70 [7] years) underwent SS-OCTA imaging (PLEX((R)) Elite 9000, Carl Zeiss Meditec, Inc., Dublin, USA). Each participant underwent 2 repeated scans of 2 scan protocols: a macular-centred 3 x 3-mm(2) and a widefield 12 x 12-mm(2) for a total of 4 acqui...
9 CitationsSource
#1Jacqueline ChuaH-Index: 17
#2Ralene SimH-Index: 3
Last. Leopold SchmettererH-Index: 81
view all 11 authors...
Diabetic retinopathy (DR) is a common complication of diabetes mellitus that disrupts the retinal microvasculature and is a leading cause of vision loss globally. Recently, optical coherence tomography angiography (OCTA) has been developed to image the retinal microvasculature, by generating 3-dimensional images based on the motion contrast of circulating blood cells. OCTA offers numerous benefits over traditional fluorescein angiography in visualizing the retinal vasculature in that it is non-i...
9 CitationsSource
#1Emily S. Levine (Tufts Medical Center)H-Index: 4
#2Malvika Arya (Tufts Medical Center)H-Index: 8
Last. Nadia K. Waheed (Tufts Medical Center)H-Index: 2
view all 9 authors...
PURPOSE: Understanding the precision of measurements on and across optical coherence tomography angiography (OCTA) devices is critical for tracking meaningful change in disease. The purpose of this study is to investigate the repeatability and reproducibility of vessel area density and vessel skeleton density measurements from various commercial OCTA devices in diabetic eyes. METHODS: Patients were imaged three consecutive times each on three different OCTA devices. En face OCTA images of the su...
5 CitationsSource