A tertiary centre experience of thoracic CT and cardiac MRI scanning in the presence of a reveal LINQ insertable cardiac monitoring system: a case series review of artefact, patient safety and data preservation.

Volume: 91, Issue: 1086, Pages: 20170615 - 20170615
Published: Jun 1, 2018
Abstract
At our tertiary cardiothoracic centre, cardiac MRI and thoracic CT scans are performed in patients with implanted LINQ devices. The degree of foreign body artefact associated with the LINQ device, and its clinical importance, has not previously been assessed. A case series review was therefore performed with a simultaneous review of patient safety and data loss events, secondary to the MRI environment.A local database search identified LINQ...