Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit

Published on Mar 18, 2005in European Radiology5.315
路 DOI :10.1007/S00330-004-2602-Z
Marcel J W Greuter21
Estimated H-index: 21
(UG: University of Groningen),
J Dorgelo6
Estimated H-index: 6
(UG: University of Groningen)
+ 1 AuthorsMatthijs Oudkerk94
Estimated H-index: 94
(UG: University of Groningen)
Sources
Abstract
Acquisition time plays a key role in the quality of cardiac multidetector computed tomography (MDCT) and is directly related to the rotation time of the scanner. The purpose of this study is to examine the influence of heart rate and a multisector reconstruction algorithm on the image quality of coronary arteries of an anthropomorphic adjustable moving heart phantom on an ECG-gated MDCT unit. The heart phantom and a coronary artery phantom were used on a MDCT unit with a rotation time of 500 ms. The movement of the heart was determined by analysis of the images taken at different phases. The results indicate that the movement of the coronary arteries on the heart phantom is comparable to that in a clinical setting. The influence of the heart rate on image quality and artifacts was determined by analysis of several heart rates between 40 and 80 bpm where the movement of the heart was synchronized using a retrospective ECG-gated acquisition protocol. The resulting reformatted volume rendering images of the moving heart and the coronary arteries were qualitatively compared as a result of the heart rate. The evaluation was performed on three independent series by two independent radiologists for the image quality of the coronary arteries and the presence of artifacts. The evaluation shows that at heart rates above 50 bpm the influence of motion artifacts in the coronary arteries becomes apparent. In addition the influence of a dedicated multisector reconstruction technique on image quality was determined. The results show that the image quality of the coronary arteries is not only related to the heart rate and that the influence of the multisector reconstruction technique becomes significant above 70 bpm. Therefore, this study proves that from the actual acquisition time per heart cycle one cannot determine an actual acquisition time, but only a mathematical acquisition time.
馃摉 Papers frequently viewed together
200011.11Radiology
References28
Newest
#1Christopher Herzog (Goethe University Frankfurt)H-Index: 23
#2Martina B. Britten (Goethe University Frankfurt)H-Index: 17
Last. Thomas J. Vogl (Goethe University Frankfurt)H-Index: 82
view all 10 authors...
The aim of this study was to investigate the accuracy of multidetector-row cardiac CT (MDCT), calcium scoring (Ca-Sc), and MDCT coronary angiography (MD CTA) in the assessment of coronary atherosclerosis. Thirty-eight patients underwent invasive coronary angiography (CA) and MDCT (collimation 4脳1 mm, pitch 1.5 mm, TI 500 ms, 120 kV, 300 mAs, and retrospective ECG-gating). Calcium scoring was calculated for the total coronary artery territory and for RCA, LCA, and LCX separately. The MD CTA serve...
Source
OBJECTIVE. We introduce stratification data for three methods (Agatston, volume, mass) obtained from one single patient population.MATERIALS AND METHODS. Measurements in 11,490 individuals scanned from 1999 to 2002 with electron-beam CT were used for this study.RESULTS. Our Agatston score ranges agree reasonably well with the Kondos values except for measurements in patients at the extreme ages, at which we sampled a wider age range and consequently had different biases of averages. Neither meth...
Source
#1Songshou Mao (UCLA Medical Center)H-Index: 25
#2Bin Lu (UCLA Medical Center)H-Index: 26
Last. Matthew J. Budoff (UCLA Medical Center)H-Index: 141
view all 5 authors...
Abstract Rationale and Objectives The R wave of the electrocardiogram is used widely as a trigger for cardiac imaging. This study was designed to determine the optimal interval between the R wave and end systole for triggering of electron-beam computed tomography (CT) in a group of patients with various heart rates who are undergoing assessment for coronary artery calcification. Materials and Methods A total of 862 consecutive asymptomatic patients referred for screening with electron-beam CT fo...
Source
#1Tobias F. Jakobs (LMU: Ludwig Maximilian University of Munich)H-Index: 39
#2Bernd J. Wintersperger (LMU: Ludwig Maximilian University of Munich)H-Index: 49
Last. Christoph R. Becker (LMU: Ludwig Maximilian University of Munich)H-Index: 96
view all 8 authors...
The aim of this study was to reduce radiation exposure in multislice CT (MSCT) coronary artery calcium screening using different tube settings, and to determinate its impact on the detection and quantification of coronary artery calcification. Forty-eight patients underwent routine MSCT coronary artery calcium scoring (Somatom VolumeZoom, Siemens, Forchheim, Germany) with retrospective ECG-gated data acquisition. Scanning was performed with a 4脳2.5-mm collimation. In each patient data acquisitio...
Source
Objective: The objective of this study was to compare the temporal resolution-related image quality of electrocardiography-gated images acquired with two multidetector computed tomography (CT) units with a moving heart phantom, at similar fixed heart rates, using half-scan and multisector acquisition modes. Methods: An adjustable moving heart phantom (Limbsandthings. Horfield, Bristol, UK) was used. Specific heart rates (47, 55, 64, 66, 69, and 73 beats per minute [bpm]) were chosen. On a Genera...
Source
#1Stefan Ulzheimer (FAU: University of Erlangen-Nuremberg)H-Index: 23
#2Willi A. Kalender (FAU: University of Erlangen-Nuremberg)H-Index: 100
Electron beam tomography (EBT) has been used for cardiac diagnosis and the quantitative assessment of coronary calcium since the late 1980s. The introduction of mechanical multi-slice spiral CT (MSCT) scanners with shorter rotation times opened new possibilities of cardiac imaging with conventional CT scanners. The purpose of this work was to qualitatively and quantitatively evaluate the performance for EBT and MSCT for the task of coronary artery calcium imaging as a function of acquisition pro...
Source
#1Sandra S. Halliburton (Cleveland Clinic)H-Index: 31
#2Arthur E. Stillman (Cleveland Clinic)H-Index: 41
Last. Richard D. White (Cleveland Clinic)H-Index: 64
view all 10 authors...
To evaluate segmented reconstruction algorithms for spiral multi-slice computed tomography (MSCT) that use data from two cardiac cycles to improve temporal resolution (蟿) for imaging of the heart.
Source
#1Koen Nieman (EUR: Erasmus University Rotterdam)H-Index: 64
#2Filippo Cademartiri (EUR: Erasmus University Rotterdam)H-Index: 80
Last. Pim J. de Feyter (EUR: Erasmus University Rotterdam)H-Index: 110
view all 6 authors...
Background鈥 Multislice spiral computed tomography (MSCT) is a promising technique for noninvasive coronary angiography, although clinical application has remained limited because of frequently incomplete interpretability, caused by motion artifacts and calcifications. Methods and Results鈥 In 59 patients (53 male, aged 58卤12 years) with suspected obstructive coronary artery disease, ECG-gated MSCT angiography was performed with a 16-slice MSCT scanner (0.42-s rotation time, 12脳0.75-mm detector co...
Source
#1T. GieslerH-Index: 7
#2Ulrich BaumH-Index: 31
Last. Stephan AchenbachH-Index: 112
view all 10 authors...
OBJECTIVE. Although multidetector CT (MDCT) with retrospectively ECG-gated image reconstruction has been shown to permit noninvasive visualization of the coronary arteries, the 125-250 msec required for image acquisition frequently causes motion artifacts. We investigated the influence of a patient's heart rate on the presence of motion artifacts and on accuracy of stenosis detection on contrast-enhanced MDCT.MATERIALS AND METHODS. In 100 patients, MDCT was performed, and ECG-gated cross-section...
Source
#1Stephan Wicky (Harvard University)H-Index: 23
#2Michael S. RosolH-Index: 7
Last. Thomas J. BradyH-Index: 102
view all 7 authors...
Purpose: We evaluated independently retrospective half scan and multisector mode manufacturer's protocols and compared them with modified acquisition protocols to determine optimal imaging parameters for cardiac scanning.Materials and Methods: Data were acquired using two fabricated gated moving phantoms. In half scan mode, the manufacturer's recommended pitch values were compared with adjacent values at different motion rates. In multisector mode, the manufacturer's protocols were compared with...
Source
Cited By32
Newest
#1Konstantin Bolwin (European Institute)H-Index: 3
#2Bj枚rn Czekalla (European Institute)H-Index: 1
Last. Klaus P. Sch盲fers (WWU: University of M眉nster)H-Index: 33
view all 5 authors...
Patient motion during medical imaging using techniques such as computed tomography (CT), magnetic resonance imaging (MRI), positron emission tomography (PET), or single emission computed tomography (SPECT) is well known to degrade images, leading to blurring effects or severe artifacts. Motion correction methods try to overcome these degrading effects. However, they need to be validated under realistic conditions. In this work, a sophisticated anthropomorphic thorax phantom is presented that com...
Source
#1H. Zou (CQMU: Chongqing Medical University)H-Index: 1
#2Y. Zhang (CQMU: Chongqing Medical University)H-Index: 1
Last. Z. Liu (CQMU: Chongqing Medical University)H-Index: 1
view all 4 authors...
Background The diagnostic accuracy of multidetector computed tomography (MDCT) in the detection of left atrial/left atrial appendage (LA/LAA) thrombus had been tested in many studies; however, the results are controversial. Aim The aim of the present study was to evaluate synthetically the diagnostic accuracy of MDCT in LA/LAA thrombus. Methods PubMed and Embase were searched for published prospective studies which compared computed tomography with transoesophageal echocardiography (TEE) in dete...
Source
#1James Otton (UNSW: University of New South Wales)H-Index: 14
#2Justin Phan (UNSW: University of New South Wales)H-Index: 5
Last. Jane McCrohon (UNSW: University of New South Wales)H-Index: 21
view all 6 authors...
Background Aggressive dose reduction strategies for cardiac CT require the prospective selection of limited cardiac phases. At lower heart rates, the period of mid-diastole is typically selected for image acquisition. We aimed to identify the effect of heart rate on the optimal CT acquisition phase within the period of mid-diastole.
Source
#1Thomas BoltzH-Index: 1
#1Almar KleinH-Index: 7
Last. Cornelis H. SlumpH-Index: 27
view all 7 authors...
Endovascular aortic repair (EVAR) is a technique which uses stent grafts to treat aortic aneurysms in patients at risk of aneurysm rupture. Although this technique has been shown to be very successful on the short term, the long term results are less optimistic due to failure of the stent graft. The pulsating blood flow applies stresses and forces to the stent graft, which can cause problems such as breakage, leakage, and migration. Therefore it is of importance to gain more insight into the in ...
Source
#1Ali Hassan (Wexham Park Hospital)H-Index: 1
#2Sarfraz Ahmed Nazir (John Radcliffe Hospital)H-Index: 5
Last. Hatem Alkadhi (UZH: University of Zurich)H-Index: 86
view all 3 authors...
Rapid advancements in multidetector row computed tomography (MDCT) are beginning to revolutionise cardiac imaging applications. As a consequence, coronary CT angiography (CTA) is fast emerging as a highly effective, noninvasive imaging technique for the assessment of coronary artery disease (CAD). Technology is improving at a robust pace, which brings with it the benefits of superior spatial and temporal resolution as well as fast volume coverage, achieved through the development of systems with...
Source
#1Thomas Boltz (Mayo Clinic)H-Index: 5
#2William Pavlicek (Mayo Clinic)H-Index: 26
Last. Metin Akay (ASU: Arizona State University)H-Index: 36
view all 6 authors...
The current work describes an anthropomorphic beating heart phantom constructed as a tool for the assessment of technological advances in cardiac x-ray computed tomography (CT). The phantom is comprised of a thorax, a compressor system, an ECG system, a beating heart with tortuous coronary arteries, and the option to add or remove pathologies such as aberrant beats, stents, and plaques. Initial trials with the phantom have shown its utility to assess temporal resolution, spatial resolution, radi...
Source
#1Wisnumurti Kristanto (UMCG: University Medical Center Groningen)H-Index: 6
#2Peter M. A. van Ooijen (UMCG: University Medical Center Groningen)H-Index: 23
Last. Matthijs Oudkerk (UMCG: University Medical Center Groningen)H-Index: 94
view all 6 authors...
Multi detector-row CT (MDCT), the current preferred method for coronary artery disease assessment, is still affected by motion artefacts. To rule out motion artefacts, qualitative image analysis is usually performed. Our study aimed to develop a quantitative image analysis for motion artefacts detection as an added value to the qualitative analysis. An anthropomorphic moving heart phantom with adjustable heart-rate was scanned on 64-MDCT and dual-source-CT. A new software technique was developed...
Source
Jan 1, 2010 in EMBC (International Conference of the IEEE Engineering in Medicine and Biology Society)
#5Limin Luo (SEU: Southeast University)H-Index: 12
This paper describes a method for the characterization of coronary artery motion using multislice computed tomography (MSCT) volume sequences. Coronary trees are first extracted by a spatial vessel tracking method in each volume of MSCT sequence. A point-based matching algorithm, with feature landmarks constraint, is then applied to match the 3-D extracted centerlines between two consecutive instants over a complete cardiac cycle. The transformation functions and correspondence matrices are esti...
Source
This website uses cookies.
We use cookies to improve your online experience. By continuing to use our website we assume you agree to the placement of these cookies.
To learn more, you can find in our Privacy Policy.