Radiation Dose to Critical Organs During Petrous Tomography

Volume: 94, Issue: 3, Pages: 623 - 627
Published: Mar 1, 1970
Abstract
It was found that during “routine” petrous bone tomography a dose of 10 R was delivered to the cornea. Simple lead shields placed over the cornea reduced this dose to about 1.0 R. The shielding did not interfere with film quality and was easily accomplished. Tables give the doses to 20 organs and/or areas in Rando phantoms with and without the lead shield and in 10 critical organs in 6 patients with the shields in...
Paper Details
Title
Radiation Dose to Critical Organs During Petrous Tomography
Published Date
Mar 1, 1970
Journal
Volume
94
Issue
3
Pages
623 - 627
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.