How MRI Compatible is “MRI Compatible”? A Systematic Comparison of Artifacts Caused by Biopsy Needles at 3.0 and 1.5 T

Volume: 36, Issue: 6, Pages: 1646 - 1657
Published: Apr 12, 2013
Abstract
This study was designed to systematically investigate artifacts caused by interventional needles recommended for use in MRI, with focus on field strength, needle/mandrin type, orientation and sequence. Eight different MRI compatible needles were placed in porcine tissue and examined at 1.5 and 3.0 T with balanced-steady-state-free-precession (B-SSFP) and T1-weighted-spoiled-gradient-echo (T1-SPGR) sequences in different orientations to B0....
Paper Details
Title
How MRI Compatible is “MRI Compatible”? A Systematic Comparison of Artifacts Caused by Biopsy Needles at 3.0 and 1.5 T
Published Date
Apr 12, 2013
Volume
36
Issue
6
Pages
1646 - 1657