Staff dose reduction during coronary angiography using low framing speed

Volume: 69, Issue: 825, Pages: 860 - 864
Published: Sep 1, 1996
Abstract
Radiation exposure of cardiac catheterization laboratory staff is a known occupational hazard, and efforts are made to obtain dose levels which are "as low as reasonably achievable". This study assessed the reduction in staff radiation exposure using cine framing at 12.5 f s-1 during coronary cineangiography, instead of 25 f s-1. Thermoluminescent dosemeters were used to measure equivalent dose at several sites for the operator, nurse assistant,...
Paper Details
Title
Staff dose reduction during coronary angiography using low framing speed
Published Date
Sep 1, 1996
Volume
69
Issue
825
Pages
860 - 864
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.