Efficacy of the Flow Re-direction Endoluminal Device for cerebral aneurysms and causes of failed deployment

Volume: 64, Issue: 6, Pages: 1213 - 1219
Published: Nov 13, 2021
Abstract
The Flow Re-direction Endoluminal Device (FRED) has recently become available for flow diversion in Japan. We have encountered cases that failed to deploy the FRED. In this study, we report our initial experience with the FRED for cerebral aneurysms and clarify the causes of failed FRED deployment.A retrospective data analysis was performed to identify patients treated with the FRED between June 2020 and March 2021. Follow-up digital subtraction...