Assessment of frequency drift on CEST MRI and dynamic correction: application to gagCEST at 7 T

Published on Jan 1, 2019in Magnetic Resonance in Medicine3.635
· DOI :10.1002/MRM.27367
Johannes Windschuh15
Estimated H-index: 15
(NYU: New York University),
Moritz Zaiss28
Estimated H-index: 28
(MPG: Max Planck Society)
+ 3 AuthorsRavinder R. Regatte41
Estimated H-index: 41
(NYU: New York University)
Sources
Abstract
PURPOSE: To investigate the effect of a frequency drift of the static magnetic field on 3D CEST MRI based on glycosaminoglycans (GAGs) of articular cartilage at 7 T and to introduce a retrospective correction method that uses the phase images of the gradient-echo readout. METHODS: Repeated gagCEST and B0 measurements were performed in a glucose model solution and in vivo in the knee joint of 3 healthy volunteers at 7 T. Phase images of the modified 3D rectangular spiral centric-reordered gradient-echo CEST sequence were used to quantify and compensate the apparent frequency drift in repeated gagCEST measurements. RESULTS: The frequency drift of the MRI scanner strongly influences the gagCEST signal in the articular cartilage of the human knee joint. The gagCEST signal in the articular cartilage is changed by 0.18%/Hz while an average drift of 0.7 ± 0.2 Hz/minute was observed. The proposed correction method can be applied retrospectively without the need of additional measurements and provides improved comparability and reproducibility for gagCEST studies. This correction method may also be of interest for other applications of CEST MRI. CONCLUSION: Prospective or retrospective correction of the frequency drift of the MRI scanner is essential for reproducible gagCEST measurements. The proposed retrospective correction method fulfills this requirement without the need of additional measurements.
📖 Papers frequently viewed together
182 Citations
8 Authors (Kejia Cai, ..., Ravinder Reddy)
418 Citations
434 Citations
References43
Newest
#1Moritz Zaiss (MPG: Max Planck Society)H-Index: 28
#2Philipp Ehses (MPG: Max Planck Society)H-Index: 17
Last. Klaus Scheffler (MPG: Max Planck Society)H-Index: 76
view all 3 authors...
: Gradient echo (GRE)-based acquisition provides a robust readout method for chemical exchange saturation transfer (CEST) at ultrahigh field (UHF). To develop a snapshot-CEST approach, the transient GRE signal and point spread function were investigated in detail, leading to optimized measurement parameters and reordering schemes for fast and robust volumetric CEST imaging. Simulation of the transient GRE signal was used to determine the optimal sequence parameters and the maximum feasible numbe...
43 CitationsSource
#1Peter C.M. van Zijl (ICSCI: Kennedy Krieger Institute)H-Index: 109
#2Wilfred W. Lam (Sunnybrook Research Institute)H-Index: 12
Last. Greg J. Stanisz (Medical University of Lublin)H-Index: 41
view all 5 authors...
Abstract Magnetization Transfer Contrast (MTC) and Chemical Exchange Saturation Transfer (CEST) experiments measure the transfer of magnetization from molecular protons to the solvent water protons, an effect that becomes apparent as an MRI signal loss (“saturation”). This allows molecular information to be accessed with the enhanced sensitivity of MRI. In analogy to Magnetic Resonance Spectroscopy (MRS), these saturation data are presented as a function of the chemical shift of participating pr...
109 CitationsSource
#1Osamu Togao (Kyushu University)H-Index: 27
#2Jochen Keupp (Philips)H-Index: 19
Last. Hiroshi Honda (Kyushu University)H-Index: 79
view all 7 authors...
Purpose To assess the quantitative performance of three-dimensional (3D) fast spin-echo (FSE) Dixon amide proton transfer (APT) imaging of brain tumors compared with B0 correction with separate mapping methods. Methods Twenty-two patients with brain tumors (54.2 ± 18.7 years old, 12 males and 10 females) were scanned at 3 Tesla (T). Z-spectra were obtained at seven different frequency offsets at ±3.1 ppm, ± 3.5 ppm, ± 3.9 ppm, and −1560 ppm. The scan was repeated three times at +3.5 ppm with ech...
28 CitationsSource
#1Guruprasad Krishnamoorthy (TU/e: Eindhoven University of Technology)H-Index: 3
#2Ravi Prakash Reddy Nanga (UPenn: University of Pennsylvania)H-Index: 24
Last. Ravinder Reddy (UPenn: University of Pennsylvania)H-Index: 66
view all 5 authors...
Purpose To develop a new faster and higher quality three-dimensional (3D) gagCEST MRI technique for reliable quantification of glycosaminoglycan (GAG) present in the human knee cartilages. Methods A new magnetization-prepared 3D gradient echo-based MRI pulse sequence has been designed to obtain the B0 inhomogeneity, B1 inhomogeneity, and CEST Z-spectra images. Results The gagCEST values of different compartments of knee cartilage are calculated using a newly developed technique for healthy subje...
36 CitationsSource
#1Feliks KoganH-Index: 18
#2Brian A. HargreavesH-Index: 44
Last. Garry E. GoldH-Index: 73
view all 3 authors...
Purpose To develop and optimize a multislice glycosaminoglycan (GAG) chemical exchange saturation transfer (GagCEST) sequence for volumetric imaging of articular cartilage, and to validate the sequence against T1ρ relaxation times in whole joint imaging of tibiotalar cartilage. Methods Ex vivo experiments were used to observe the effect of the number of partitions and shot TR on signal-to-noise ratio and measured GagCESTasym. GagCEST imaging of the entire tibiotalar joint was also performed on 1...
30 CitationsSource
#1Patrick Schuenke (DKFZ: German Cancer Research Center)H-Index: 9
#2Johannes Windschuh (DKFZ: German Cancer Research Center)H-Index: 15
Last. Moritz Zaiss (DKFZ: German Cancer Research Center)H-Index: 28
view all 6 authors...
Purpose Together with the development of MRI contrasts that are inherently small in their magnitude, increased magnetic field accuracy is also required. Hence, mapping of the static magnetic field (B0) and the excitation field (B1) is not only important to feedback shim algorithms, but also for postprocess contrast-correction procedures. Methods A novel field-inhomogeneity mapping method is presented that allows simultaneous mapping of the water shift and B1 (WASABI) using an off-resonant rectan...
57 CitationsSource
#1David R. Roalf (UPenn: University of Pennsylvania)H-Index: 44
#2Ravi Prakash Reddy Nanga (UPenn: University of Pennsylvania)H-Index: 24
Last. Bruce I. Turetsky (UPenn: University of Pennsylvania)H-Index: 64
view all 14 authors...
Glutamate imaging (GluCEST) reveals lower brain GluCEST contrast in patients on the psychosis spectrum
50 CitationsSource
#1Sjoerd B. VosH-Index: 21
#2Chantal M. W. Tax (UU: Utrecht University)H-Index: 17
Last. Martijn Froeling (UU: Utrecht University)H-Index: 23
view all 6 authors...
Purpose To investigate previously unreported effects of signal drift as a result of temporal scanner instability on diffusion MRI data analysis and to propose a method to correct this signal drift. Methods We investigated the signal magnitude of non-diffusion-weighted EPI volumes in a series of diffusion-weighted imaging experiments to determine whether signal magnitude changes over time. Different scan protocols and scanners from multiple vendors were used to verify this on phantom data, and th...
101 CitationsSource
#1Jihong WangH-Index: 15
#2Joseph Weygand (University of Texas MD Anderson Cancer Center)H-Index: 5
Last. Jinyuan Zhou (Johns Hopkins University)H-Index: 1
view all 9 authors...
Magnetic Resonance Imaging of Glucose Uptake and Metabolism in Patients with Head and Neck Cancer
44 CitationsSource
#1Anja Müller-LutzH-Index: 13
#2Felix MatuschkeH-Index: 5
Last. Hans-Jörg WittsackH-Index: 35
view all 7 authors...
Abstract Purpose To optimize B 0 -field inhomogeneity correction for chemical exchange saturation transfer (CEST) imaging by investigating different water saturation shift referencing (WASSR) Z-spectrum shapes and different frequency correction techniques. Methods WASSR Z-spectra were simulated for different B 1 -fields and pulse durations (PD). Two parameter settings were used for further simulations and experiments (WASSR1: B 1 = 0.1 μT, PD = 50 ms; WASSR2: B 1 = 0.3 μT, PD = 40 ms). Four freq...
6 CitationsSource
Cited By24
Newest
#1Anina Seidemo (Lund University)H-Index: 2
#2Patrick M. Lehmann (Lund University)
Last. Linda Knutsson (JHUSOM: Johns Hopkins University School of Medicine)H-Index: 20
view all 10 authors...
Dynamic glucose-enhanced (DGE) magnetic resonance imaging (MRI) has shown potential for tumor imaging using D-glucose as a biodegradable contrast agent. The DGE signal change is small at 3 T (around 1%) and accurate detection is hampered by motion. The intravenous D-glucose injection is associated with transient side effects that can indirectly generate subject movements. In this study, the aim was to study DGE arterial input functions (AIFs) in healthy volunteers at 3 T for different scanning p...
Source
#1Ran Sui (Johns Hopkins University)H-Index: 1
#3Yuguo Li (ICSCI: Kennedy Krieger Institute)H-Index: 13
Last. Jiadi Xu (ICSCI: Kennedy Krieger Institute)H-Index: 34
view all 8 authors...
PURPOSE To develop a steady-state saturation with radial readout chemical exchange saturation transfer (starCEST) for acquiring CEST images at 3 Tesla (T). The polynomial Lorentzian line-shape fitting approach was further developed for extracting amideCEST intensities at this field. METHOD StarCEST MRI using periodically rotated overlapping parallel lines with enhanced reconstruction-based spatial sampling was implemented to acquire Z-spectra that are robust to brain motion. Multi-linear singula...
1 CitationsSource
#2Ulrich PilatusH-Index: 30
Last. Manoj Shrestha (Goethe University Frankfurt)H-Index: 4
view all 5 authors...
In this study, we present a CEST-EPI sequence for rapid multislice data acquisition, which records multiple slices and adjacent frequency offsets after reaching steady-state magnetization once. The time saving compared with an already modified conventional multislice acquisition scheme, which rebuilds the steady state repeatedly during a frequency offset change, amounts to almost 50%.
Source
#1Soo Hyun Shin (University of California, Berkeley)H-Index: 1
#2Michael F. Wendland (University of California, Berkeley)H-Index: 66
Last. Moriel H. Vandsburger (University of California, Berkeley)H-Index: 13
view all 3 authors...
Purpose We demonstrate a method of delayed urea differential enhancement CEST for probing urea recycling action of the kidney using expanded multi-pool Lorentzian fitting and apparent exchange-dependent relaxation compensation. Methods T1 correction of urea CEST contrast by apparent exchange-dependent relaxation was tested in phantoms. Nine mice were scanned at 7 Tesla following intraperitoneal injection of 2M 150 μL urea, and later saline. T1 maps and Z-spectra were acquired before and 20 and 4...
Source
PURPOSE To explore the impact of temporal motion-induced coil sensitivity changes on CEST-MRI at 7T and its correction using interleaved volumetric EPI navigators, which are applied for real-time motion correction. METHODS Five healthy volunteers were scanned via CEST. A 4-fold correction pipeline allowed the mitigation of (1) motion, (2) motion-induced coil sensitivity variations, ΔB1- , (3) motion-induced static magnetic field inhomogeneities, ΔB0 , and (4) spatially varying transmit RF field ...
Source
#1Ruibin Liu (ZJU: Zhejiang University)H-Index: 2
Last. Yi Zhang (ZJU: Zhejiang University)H-Index: 20
view all 8 authors...
PURPOSE To correct the temporal B0 drift in chemical exchange saturation transfer (CEST) imaging in real-time with extra free-induction-decay (FID) readout. THEORY AND METHODS The frequency stabilization module of the recently proposed frequency-stabilized CEST (FS-CEST) sequence was further simplified by replacing the original three k-space lines of gradient-echo (GRE) readout with a single k-space line of FID readout. The B0 drift was quantified using the phase difference between the odd and e...
2 CitationsSource
#1S Mueller (MPG: Max Planck Society)H-Index: 1
#2Klaus Scheffler (MPG: Max Planck Society)H-Index: 76
Last. Moritz Zaiss (MPG: Max Planck Society)H-Index: 28
view all 3 authors...
Chemical exchange saturation transfer (CEST) MRI is currently set to become part of clinical routine as it enables indirect detection of low concentrated molecules and proteins. Recently, intermediate to fast exchanging functional groups of glucose and its derivatives, glutamate and dextran, have gained attention as promising CEST contrast agents. To increase the specificity of CEST MRI for certain functional groups, the presaturation module is commonly optimized. At an early stage, this is perf...
1 CitationsSource
#2Suyi Cao (JHUSOM: Johns Hopkins University School of Medicine)H-Index: 3
#3Raymond C. Koehler (JHUSOM: Johns Hopkins University School of Medicine)H-Index: 73
Last. Jiadi Xu (JHUSOM: Johns Hopkins University School of Medicine)H-Index: 34
view all 5 authors...
PURPOSE To obtain high-sensitivity CEST maps by exploiting the spatiotemporal correlation between CEST images. METHODS A postprocessing method accomplished by multilinear singular value decomposition (MLSVD) was used to enhance the CEST SNR by exploiting the correlation between the Z-spectrum for each voxel and the low-rank property of the overall CEST data. The performance of this method was evaluated using CrCEST in ischemic mouse brain at 11.7 tesla. Then, MLSVD CEST was applied to obtain Cr,...
4 CitationsSource
#1Lebina S. Kakkar (UCL: University College London)H-Index: 2
#2Muhammad Usman (UCL: University College London)H-Index: 68
Last. David Atkinson (UCL: University College London)H-Index: 68
view all 5 authors...
Multi-parametric MRI is increasingly used for prostate cancer detection. Improving information from current sequences, such as T2-weighted and diffusion-weighted (DW) imaging, and additional sequences, such as magnetic resonance spectroscopy (MRS) and chemical exchange saturation transfer (CEST), may enhance the performance of multi-parametric MRI. The majority of these techniques are sensitive to B0-field variations and may result in image distortions including signal pile-up and stretching (ec...
Source
#1S Mueller (MPG: Max Planck Society)H-Index: 1
#2Rüdiger Stirnberg (German Center for Neurodegenerative Diseases)H-Index: 12
Last. Moritz Zaiss (MPG: Max Planck Society)H-Index: 28
view all 7 authors...
PURPOSE: CEST MRI enables imaging of distributions of low-concentrated metabolites as well as proteins and peptides and their alterations in diseases. CEST examinations often suffer from low spatial resolution, long acquisition times, and concomitant motion artifacts. This work aims to maximize both resolution and volume coverage with a 3D-EPI snapshot CEST approach at 3T, allowing for fast and robust whole-brain CEST MRI. METHODS: Resolution and temporal SNR of 3D-EPI examinations with nonselec...
6 CitationsSource