Magnetic field–induced interactions between phones containing magnets and cardiovascular implantable electronic devices: Flip it to be safe?

Volume: 19, Issue: 3, Pages: 372 - 380
Published: Mar 1, 2022
Abstract
Recent case reports and small studies have reported activation of the magnet-sensitive switches in cardiovascular implantable electronic devices (CIEDs) by the new iPhone 12 series, initiating asynchronous pacing in pacemakers and suspension of antitachycardia therapies in implantable cardioverter-defibrillators (ICDs).The purpose of this prospective single-center observational study was to quantify the risk of magnetic field interactions of the...
Paper Details
Title
Magnetic field–induced interactions between phones containing magnets and cardiovascular implantable electronic devices: Flip it to be safe?
Published Date
Mar 1, 2022
Volume
19
Issue
3
Pages
372 - 380
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.