Optical emission‐based phantom to verify coincidence of radiotherapy and imaging isocenters on an MR‐linac

Volume: 22, Issue: 9, Pages: 252 - 261
Published: Aug 19, 2021
Abstract
Purpose Demonstrate a novel phantom design using a remote camera imaging method capable of concurrently measuring the position of the x‐ray isocenter and the magnetic resonance imaging (MRI) isocenter on an MR‐linac. Methods A conical frustum with distinct geometric features was machined out of plastic. The phantom was submerged in a small water tank, and aligned using room lasers on a MRIdian MR‐linac (ViewRay Inc., Cleveland, OH). The phantom...
Paper Details
Title
Optical emission‐based phantom to verify coincidence of radiotherapy and imaging isocenters on an MR‐linac
Published Date
Aug 19, 2021
Volume
22
Issue
9
Pages
252 - 261
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.