Preliminary Optimized Design of a High-resolution PET Detector with a 0.5 mm Crystal Size

Published: Oct 31, 2020
Abstract
The purpose of this study is to develop a PET detector with a 0.5 mm crystal size for high-resolution preclinical PET imaging. The methods to improve the decoding performances of the detectors, including the wedge-shaped light guides with different thicknesses, different methods to suppress the dark count noises in the SiPMs and different methods to calculate the flood maps, were investigated. The experimental results show that the PET detector...