Adverse device‐device interaction between pacemaker and subcutaneous implantable cardiac defibrillator

Volume: 44, Issue: 11, Pages: 1944 - 1948
Published: Jul 31, 2021
Abstract
A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient...
Paper Details
Title
Adverse device‐device interaction between pacemaker and subcutaneous implantable cardiac defibrillator
Published Date
Jul 31, 2021
Volume
44
Issue
11
Pages
1944 - 1948
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.