DOSE EVALUATION AND PROPOSAL OF LOCAL DIAGNOSTIC REFERENCE LEVELS FOR PAEDIATRIC CARDIAC CATHETERIZATIONS PERFORMED ON A HIGH-SENSITIVITY ANGIOGRAPHIC SYSTEM ALLOWING LOW-DOSE IMAGING

Volume: 195, Issue: 3-4, Pages: 279 - 288
Published: Jun 9, 2021
Abstract
Radiation doses from paediatric cardiac catheterizations were analysed based on procedure type and patient weight, and local diagnostic reference levels (LDRLs) were proposed. The procedures were performed on a Siemens Artis Q.zen biplane system with high-sensitivity detectors allowing low-dose imaging. Good radiological practice, e.g. adapting dose level and frame rate continuously and minimizing the x-ray field with collimators, was routine...
Paper Details
Title
DOSE EVALUATION AND PROPOSAL OF LOCAL DIAGNOSTIC REFERENCE LEVELS FOR PAEDIATRIC CARDIAC CATHETERIZATIONS PERFORMED ON A HIGH-SENSITIVITY ANGIOGRAPHIC SYSTEM ALLOWING LOW-DOSE IMAGING
Published Date
Jun 9, 2021
Volume
195
Issue
3-4
Pages
279 - 288
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.