Automated algorithm for calculation of setup corrections and planning target volume margins for offline image‐guided radiotherapy protocols

Volume: 22, Issue: 7, Pages: 137 - 146
Published: Jun 9, 2021
Abstract
Purpose Each radiotherapy center should have a site‐specific planning target volume (PTV) margins and image‐guided (IG) radiotherapy (IGRT) correction protocols to compensate for the geometric errors that can occur during treatment. This study developed an automated algorithm for the calculation and evaluation of these parameters from cone beam computed tomography (CBCT)‐based IG‐intensity modulated radiotherapy (IG‐IMRT) treatment. Methods and...
Paper Details
Title
Automated algorithm for calculation of setup corrections and planning target volume margins for offline image‐guided radiotherapy protocols
Published Date
Jun 9, 2021
Volume
22
Issue
7
Pages
137 - 146