Polarity status of trigger signal during ECG-gating affects parameters of LV function in gated myocardial perfusion SPECT

Volume: 29, Issue: 5, Pages: 2199 - 2209
Published: May 25, 2021
Abstract
The polarity status is one of the important items of specifications of trigger signal from cardiac trigger monitors with two options, either positive or negative. Some systems allow the user to set the polarity of trigger signal before imaging. Efforts should be made to set the polarity status according to the recommendations provided by the manufacturers. In case of inappropriate selection, changes in computation of end-systolic and...
Paper Details
Title
Polarity status of trigger signal during ECG-gating affects parameters of LV function in gated myocardial perfusion SPECT
Published Date
May 25, 2021
Volume
29
Issue
5
Pages
2199 - 2209