On the MLC leaves alignment in the direction orthogonal to movement

Volume: 22, Issue: 6, Pages: 268 - 273
Published: May 30, 2021
Abstract
The main focus of the recommended spatial accuracy tests for the multi‐leaf collimators (MLC) is calibration of the leaf position along the movement direction and overall alignment to the radiation isocenter. No explicit attention was typically paid to the alignment of the leaves from the opposing banks in the direction orthogonal to movement. This paper is a case study demonstrating that verification of such alignment at the time of acceptance...
Paper Details
Title
On the MLC leaves alignment in the direction orthogonal to movement
Published Date
May 30, 2021
Volume
22
Issue
6
Pages
268 - 273
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.