Temporal and Axial Quantitative Uniformity Measurements of Total-body PET Systems

Volume: 62, Pages: 3041 - 3041
Published: May 1, 2021
Abstract
3041 null Objectives: null null The ability of positron emission tomography (PET) scanners to accurately reproduce standardized uptake values (SUV) serves as an important imaging biomarker in oncologic studies. We aimed to develop a practical quality control test to verify that an uEXPLORER total-body PET system produces accurate SUVs between and after calibrations and across the 194.0 cm axial field-of-view (AFOV). A 30 cm axial length...
Paper Details
Title
Temporal and Axial Quantitative Uniformity Measurements of Total-body PET Systems
Published Date
May 1, 2021
Volume
62
Pages
3041 - 3041
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.