Commissioning of a clinical pencil beam scanning proton therapy unit for ultra-high dose rates (FLASH)

Volume: 48, Issue: 7, Pages: 4017 - 4026
Published: Jul 1, 2021
Abstract
Purpose null The purpose of this work was to provide a flexible platform for FLASH research with protons by adapting a former clinical pencil beam scanning gantry to irradiations with ultra-high dose rates. null Methods null PSI Gantry 1​ treated patients until December 2018. We optimized the beamline parameters to transport the 250 MeV beam extracted from the PSI COMET accelerator to the treatment room, maximizing the transmission of beam...
Paper Details
Title
Commissioning of a clinical pencil beam scanning proton therapy unit for ultra-high dose rates (FLASH)
Published Date
Jul 1, 2021
Volume
48
Issue
7
Pages
4017 - 4026
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.