A patient risk model to determine the optimal output constancy check frequency for a radiotherapy machine

Volume: 84, Pages: 192 - 197
Published: Apr 1, 2021
Abstract
Purpose The output constancy check, a basic quality control (QC) item for radiotherapy machines, is performed daily according to suggestions in technical reports by experienced experts. In this study, a patient risk model was built to determine the optimal frequency of an output constancy check for a specific radiotherapy machine. Methods and materials The method was based on the patient risk model and comprised three steps: 1) the power...