A strategy to prevent a temperature‐induced MRI artifact in warm liquid phantoms due to convection currents

Volume: 34, Issue: 6
Published: Feb 14, 2021
Abstract
MRI phantom studies often fail to mimic the temperature of the human body, which can negatively impact accuracy. An artifact induced by increasing temperature in liquid phantoms was observed, presenting a significant challenge to temperature-controlled experiments. In this study we characterize and provide a solution to eliminate this temperature-induced MRI artifact. Low concentration (0.5-2.5 mM) agar phantoms were prepared. Utilizing a...
Paper Details
Title
A strategy to prevent a temperature‐induced MRI artifact in warm liquid phantoms due to convection currents
Published Date
Feb 14, 2021
Volume
34
Issue
6
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.