A strategy to prevent a temperature‐induced MRI artifact in warm liquid phantoms due to convection currents
Abstract
MRI phantom studies often fail to mimic the temperature of the human body, which can negatively impact accuracy. An artifact induced by increasing temperature in liquid phantoms was observed, presenting a significant challenge to temperature-controlled experiments. In this study we characterize and provide a solution to eliminate this temperature-induced MRI artifact. Low concentration (0.5-2.5 mM) agar phantoms were prepared. Utilizing a...