Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy
Abstract
To implement dual-energy cone-beam computed tomography (CBCT), a 1 to 3 mm thick copper filter was applied in front of the kilovolt generator mounted on a medical linear accelerator. The energy calibration was performed from three energy peaks among the electromagnetic radiation, gamma rays, emitted by the SUP192/SUPIr source. The energy spectrum was measured using a cadmium zinc telluride (CZT) detector. A figure-of-merit (FOM) value was...
Paper Details
Title
Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy
Published Date
Dec 31, 2020
Journal
Volume
25
Issue
4
Pages
614 - 619
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History