Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy

Volume: 25, Issue: 4, Pages: 614 - 619
Published: Dec 31, 2020
Abstract
To implement dual-energy cone-beam computed tomography (CBCT), a 1 to 3 mm thick copper filter was applied in front of the kilovolt generator mounted on a medical linear accelerator. The energy calibration was performed from three energy peaks among the electromagnetic radiation, gamma rays, emitted by the SUP192/SUPIr source. The energy spectrum was measured using a cadmium zinc telluride (CZT) detector. A figure-of-merit (FOM) value was...
Paper Details
Title
Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy
Published Date
Dec 31, 2020
Volume
25
Issue
4
Pages
614 - 619