Subcutaneous Air Around Receiver-Stimulator Causing Open-Circuit Failures: An Uncommon But Readily Treatable Condition Related to CPAP Use After Cochlear Implant Surgery

Volume: 42, Issue: 2, Pages: e157 - e160
Published: Dec 8, 2020
Abstract
Objective: To raise awareness of a unique complication associated with continuous positive airway pressure (CPAP) use after cochlear implantation to improve early detection, prevent unnecessary testing, and facilitate treatment. Study Design: Case series. Setting: Tertiary referral center. Subjects: Patients who developed subcutaneous air around the receiver-stimulator device and the associated ground electrode in the setting of CPAP usage,...
Paper Details
Title
Subcutaneous Air Around Receiver-Stimulator Causing Open-Circuit Failures: An Uncommon But Readily Treatable Condition Related to CPAP Use After Cochlear Implant Surgery
Published Date
Dec 8, 2020
Volume
42
Issue
2
Pages
e157 - e160
Citation AnalysisPro