Bowstring Phenomenon in Renal Artery Aneurysm Exclusion Using a Viabahn Stent Graft

Volume: 55, Issue: 4, Pages: 402 - 404
Published: Nov 26, 2020
Abstract
A 77-year-old man presented with an incidental finding of right renal artery aneurysm without symptoms. Computed tomography revealed a 22 mm saccular aneurysm with a wide neck at the main renal artery trunk. An 8 × 100 mm Viabahn stent graft (W. L. Gore, Flagstaff, AZ) was deployed by fully pulling back the guiding sheath. However, the deployment knob was not able to be pulled. We returned the guiding sheath to the original position and...