Validation of the Kinetik Blood Pressure Monitor—Series 1 for use in adults at home and in clinical settings, according to the 2002 European Society of Hypertension International Protocol on the validation of blood pressure devices

Volume: 35, Issue: 11, Pages: 1046 - 1050
Published: Nov 22, 2020
Abstract
The aim of this study was to assess the blood pressure (BP) measurement accuracy of the Kinetik Blood Pressure Monitor—Series 1 (BPM-1) for use in home or clinical settings according to the 2002 European Society of Hypertension International Protocol (ESH-IP). Forty-two participants were recruited to fulfil the required number of systolic and diastolic BP measurements according to the ESH-IP. Nine sequential same-arm BP readings were measured...
Paper Details
Title
Validation of the Kinetik Blood Pressure Monitor—Series 1 for use in adults at home and in clinical settings, according to the 2002 European Society of Hypertension International Protocol on the validation of blood pressure devices
Published Date
Nov 22, 2020
Volume
35
Issue
11
Pages
1046 - 1050
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.