Improving Image Quality and Reducing Radiation Dose for Pediatric CT by Using Deep Learning Reconstruction

Volume: 298, Issue: 1, Pages: 180 - 188
Published: Jan 1, 2021
Abstract
Background CT deep learning reconstruction (DLR) algorithms have been developed to remove image noise. How the DLR affects image quality and radiation dose reduction has yet to be fully investigated. Purpose To investigate a DLR algorithm’s dose reduction and image quality improvement for pediatric CT. Materials and Methods DLR was compared with filtered back projection (FBP), statistical-based iterative reconstruction (SBIR), and model-based...
Paper Details
Title
Improving Image Quality and Reducing Radiation Dose for Pediatric CT by Using Deep Learning Reconstruction
Published Date
Jan 1, 2021
Journal
Volume
298
Issue
1
Pages
180 - 188
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.