A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy

Volume: 197, Issue: 3, Pages: 198 - 208
Published: Jun 19, 2020
Abstract
High-dose 6‑MV radiotherapy may cause cardiac implantable electronic devices (CIEDs) to malfunction. To assess CIED malfunctions resulting from direct exposure up to 10 Gy, 100 pacemakers (PMs) and 40 implantable cardioverter-defibrillators (ICDs) were evaluated. CIEDs underwent baseline interrogation. In ICDs, antitachycardia therapies were disabled via the programmer while the detection windows were left enabled. A computed tomography (CT)...
Paper Details
Title
A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy
Published Date
Jun 19, 2020
Volume
197
Issue
3
Pages
198 - 208
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.