The Electrode‐Tissue Interface: The Revolutionary Role of Steroid‐Elution

Volume: 37, Issue: 9, Pages: 1232 - 1249
Published: Jul 29, 2014
Abstract
The electrode-tissue interface is that area lying between the cathode of a low-voltage implantable pacemaker or cardioverter-defibrillator (ICD) lead and the endocardium or epi-myocardium of the cardiac chamber being paced. The electrical stimulus that is delivered to this interface is responsible for myocyte depolarization with consequent cardiac contraction. The process by which this occurs is reasonably well understood and any explanation...
Paper Details
Title
The Electrode‐Tissue Interface: The Revolutionary Role of Steroid‐Elution
Published Date
Jul 29, 2014
Volume
37
Issue
9
Pages
1232 - 1249