IAEA methodology for on-site end-to-end IMRT/VMAT audits: an international pilot study
Abstract
Background: The IAEA has developed and tested an on-site, end-to-end IMRT/VMAT dosimetry audit methodology for head and neck cases using an anthropomorphic phantom. The audit methodology is described, and the results of the international pilot testing are presented.Material and methods: The audit utilizes a specially designed, commercially available anthropomorphic phantom capable of accommodating a small volume ion chamber (IC) in four...
Paper Details
Title
IAEA methodology for on-site end-to-end IMRT/VMAT audits: an international pilot study
Published Date
Nov 20, 2019
Journal
Volume
59
Issue
2
Pages
141 - 148