Safe guidewire visualization using the modes of a PTx transmit array MR system

Volume: 83, Issue: 6, Pages: 2343 - 2355
Published: Nov 13, 2019
Abstract
MRI-guided cardiovascular intervention using standard metal guidewires can produce focal tissue heating caused by induced radiofrequency guidewire currents. It has been shown that safe operation is made possible by using parallel transmit radiofrequency coils driven in the null current mode, which does not induce radiofrequency currents and hence allows safe tissue visualization. We propose that the maximum current modes, usually considered...