Repeatability of Wide-field Optical Coherence Tomography Angiography in Normal Retina.

Published on May 1, 2019in Translational Vision Science & Technology2.112
· DOI :10.1167/TVST.8.3.6
Michael Eastline1
Estimated H-index: 1
(University of Bern),
Marion R. Munk12
Estimated H-index: 12
(University of Bern)
+ 5 AuthorsMartin S. Zinkernagel25
Estimated H-index: 25
(University of Bern)
Sources
Abstract
Purpose: We evaluated the repeatability of wide-field en face swept-source optical coherence tomography angiography (SS-OCTA) in healthy subjects. Methods: Healthy subjects underwent two imaging sessions, on average 8 days apart, with a 100 kHz SS-OCTA instrument. The imaging protocol included a central 3 × 3 and 12 × 12 mm scans of the four quadrants resulting in more than a 70° wide-field OCTA of the posterior pole. Quantitative analysis was performed using the inbuilt Macular Density Algorithm Version v0.6.1 and AngioTool software. Consistency for the foveal avascular zone (FAZ), vessel density, and perfusion density of the superficial and deep capillary plexus slabs and the wide-field OCTA superficial slab, and the number of artefacts on the wide-field images were assessed. Results: A total of 21 healthy volunteers (seven men and 14 women; mean age 32 years; range, 18-61; standard deviation, 10.28 years) were included in this analysis. Internal consistency was highest for FAZ area with an intraclass correlation (ICC) = 0.998 (95% confidence interval [CI], 0.997-0.999), a FAZ perimeter with an ICC = 0.995 (95% CI, 0.990-0.997), a FAZ circularity with an ICC= 0.976 (95% CI, 0.956-0.987), followed by the vessel density of the inner ring in the superficial slab with an ICC = 0.834 (95% CI, 0.691-0.911), and a vessel density of the inner ring in the deep slab with an ICC = 0.523 (95% CI, 0.113-0.744).The reproducibility of the average vessels length of the wide-field OCTA cropped images was strong (ICC = 0.801; 95% CI, 0.624-0.895), followed by the reproducibility of total number of junctions (ICC = 0.795; 95% CI, 0.613-0.892) and the vessels percentage area (ICC = 0.662; 95% CI, 0.361-0.821). Conclusions: The level of reproducibility for assessing the microvascular anatomy in wide-field OCTA is strong and can be used to quantify microvascular changes over time. Refinements in analysis strategies and a consensus of which parameters are most useful for quantitative assessment of wide-field OCTA images would be useful in the future. Translational Relevance: These findings bridge the gap between basic imaging research and clinical use for quantitative wide-field OCTA.
📖 Papers frequently viewed together
10 Citations
4 Citations
26 Citations
References12
Newest
#1Fred K. Chen (UWA: University of Western Australia)H-Index: 29
#2Moreno Menghini (UWA: University of Western Australia)H-Index: 10
Last. Danuta M. Sampson (UWA: University of Western Australia)H-Index: 4
view all 6 authors...
Purpose: To measure intrasession repeatability and interocular symmetry of the foveal avascular zone area (FAZA) and superficial retinal vessel density (SRVD) using AngioVue Analytics optical coherence tomography angiography (OCTA). Methods: Fifty healthy individuals were prospectively enrolled. OCTA scans (3 × 3 and 6 × 6 mm) were acquired twice in right and once in left eyes. FAZA (with and without rescaling) and SRVD for 18 regions (whole, fovea, parafovea, six parafoveal subregions, and nine...
20 CitationsSource
#1Chris Or (Tufts Medical Center)H-Index: 4
#2Almyr S Sabrosa (Tufts Medical Center)H-Index: 4
Last. Nadia K Waheed (Tufts Medical Center)H-Index: 12
view all 5 authors...
: As ischemia remains a key prognostic factor in the management of various diseases including diabetic retinopathy, an increasing amount of research has been dedicated to its quantification as a potential biomarker. Advancements in the quantification of retinal ischemia have been made with the imaging modalities of fluorescein angiography (FA), ultra-widefield imaging (UWF), and optical coherence tomography angiography (OCTA), with each imaging modality offering certain benefits over the others....
17 CitationsSource
#1Karen B. Schaal (University of Bern)H-Index: 13
#2Marion R. MunkH-Index: 19
Last. Sebastian WolfH-Index: 71
view all 6 authors...
PURPOSE To detect vascular abnormalities in diabetic retinopathy using swept-source optical coherence tomography angiography (SS-OCTA) widefield images, and to compare the findings with color fundus photographs (CFPs) using Early Treatment Diabetic Retinopathy Study severity grading. METHODS 3 mm × 3 mm and 12 mm × 12 mm scans were acquired to cover 70° to 80° of the posterior pole using a 100-kHz SS-OCTA instrument. Two masked graders assessed the presence of vascular abnormalities on SS-OCTA a...
45 CitationsSource
#1Jianqin Lei (Xi'an Jiaotong University)H-Index: 13
#2Mary K Durbin (ZEISS: Carl Zeiss AG)H-Index: 37
Last. Srinivas R. Sadda (UCLA: University of California, Los Angeles)H-Index: 79
view all 8 authors...
Importance The repeatability and reproducibility of quantitative metrics from optical coherence tomographic angiography (OCTA) must be assessed before these data can be confidently interpreted in clinical research and practice. Objective To evaluate the repeatability and reproducibility of OCTA-derived retinal vascular quantitative metrics. Design, Setting and Participants In this cross-sectional study, 21 healthy volunteers (42 eyes) and 22 patients with retinal disease (22 eyes), including 14 ...
100 CitationsSource
#1Özge Yanik Odabaş (Ankara University)H-Index: 1
#2Sibel DemirelH-Index: 11
Last. Figen BatıoğluH-Index: 17
view all 4 authors...
Purpose:To evaluate the repeatability of vessel density and superficial and deep foveal avascular zone measurements using optical coherence tomography angiography, and to specify a diurnal change range.Methods:Forty-six eyes of 25 healthy individuals were included. Optical coherence tomography angio
26 CitationsSource
#1Luigi De Vitis (UniSR: Vita-Salute San Raffaele University)H-Index: 7
#2Lucia Benatti (UniSR: Vita-Salute San Raffaele University)H-Index: 6
Last. Francesco Bandello (UniSR: Vita-Salute San Raffaele University)H-Index: 64
view all 8 authors...
Purpose: To compare the performance of two different spectral-domain optical coherence tomography angiography (OCTA) devices in clinical practice by evaluating examination execution time and the number of motion artifacts per image. Methods: Seventy-six patients affected by different ocular diseases and 13 healthy subjects consecutively underwent assessments by two different OCTA devices: AngioPlex (Zeiss Meditec, Inc., Dublin, Calif., USA) and AngioVue (OptoVue, Inc., Fremont, Calif., USA). Two...
16 CitationsSource
#1Abtin Shahlaee (Thomas Jefferson University)H-Index: 15
#2Maria Pefkianaki (Thomas Jefferson University)H-Index: 2
Last. Allen C. Ho (Thomas Jefferson University)H-Index: 44
view all 4 authors...
Purpose To measure foveal avascular zone (FAZ) dimensions in healthy eyes using optical coherence tomography angiography (OCTA) and calculate interobserver variability. Design Reliability analysis. Methods Thirty-four eyes of 17 healthy subjects underwent OCTA at the Retina Service of Wills Eye Hospital. Two masked graders performed measurements of FAZ dimensions including area, perimeter, and maximum horizontal and vertical diameters using ImageJ. Intraclass correlation coefficient (ICC) betwee...
125 CitationsSource
#1Yali Jia (OHSU: Oregon Health & Science University)H-Index: 43
#2Steven T. Bailey (OHSU: Oregon Health & Science University)H-Index: 24
Last. David Huang (OHSU: Oregon Health & Science University)H-Index: 93
view all 12 authors...
Purpose To detect and quantify choroidal neovascularization (CNV) in patients with age-related macular degeneration (AMD) using optical coherence tomography (OCT) angiography. Design Observational, cross-sectional study. Participants A total of 5 normal subjects and 5 subjects with neovascular AMD were included. Methods A total of 5 eyes with neovascular AMD and 5 normal age-matched controls were scanned by a high-speed (100 000 A-scans/seconds) 1050-nm wavelength swept-source OCT. The macular a...
518 CitationsSource
#1Yuuki Watanabe (YU: Yamagata University)H-Index: 16
#2Yuhei Takahashi (YU: Yamagata University)H-Index: 1
Last. Hiroshi Numazawa (YU: Yamagata University)H-Index: 2
view all 3 authors...
We demonstrate intensity-based optical coherence tomography (OCT) angiography using the squared difference of two sequential frames with bulk-tissue-motion (BTM) correction. This motion correction was performed by minimization of the sum of the pixel values using axial- and lateral-pixel–shifted structural OCT images. We extract the BTM-corrected image from a total of 25 calculated OCT angiographic images. Image processing was accelerated by a graphics processing unit (GPU) with many stream proc...
21 CitationsSource
#1Enrique ZudaireH-Index: 27
#2Laure Gambardella (Babraham Institute)H-Index: 20
Last. Sonja Vermeren (Babraham Institute)H-Index: 13
view all 4 authors...
Angiogenesis is the generation of mature vascular networks from pre-existing vessels. Angiogenesis is crucial during the organism' development, for wound healing and for the female reproductive cycle. Several murine experimental systems are well suited for studying developmental and pathological angiogenesis. They include the embryonic hindbrain, the post-natal retina and allantois explants. In these systems vascular networks are visualised by appropriate staining procedures followed by microsco...
395 CitationsSource
Cited By14
Newest
#1Ella Courtie (University of Birmingham)H-Index: 1
#2Aditya Kale (University Hospitals Birmingham NHS Foundation Trust)H-Index: 3
Last. Alastair K Denniston (University Hospitals Birmingham NHS Foundation Trust)H-Index: 31
view all 0 authors...
To assess the stability of retinal structure and blood flow measures over time and in different clinical settings using portable optical coherence tomography angiography (OCTA) as a potential biomarker of central perfusion in critical illness, 18 oesophagectomy patients completed retinal structure and blood flow measurements by portable OCT and OCTA in the eye clinic and intensive therapy unit (ITU) across three timepoints: (1) pre-operation in a clinic setting; (2) 24–48 h post-operation during...
Source
#1Pasha Anvari (IUMS: Iran University of Medical Sciences)H-Index: 6
#2Maryam Ashrafkhorasani (IUMS: Iran University of Medical Sciences)
Last. Khalil Ghasemi Falavarjani (IUMS: Iran University of Medical Sciences)H-Index: 30
view all 4 authors...
We performed a comprehensive search of the published literature in PubMed and Google Scholar to identify types, prevalence, etiology, clinical impact, and current methods for correction of various artifacts in optical coherence tomography angiography (OCTA) images. We found that the prevalence of OCTA image artifacts is fairly high. Artifacts associated with eye motion, misidentification of retinal layers, projections, and low optical coherence tomography signal are the most prevalent types. Art...
Source
#1Thibaud MathisH-Index: 9
#2Sarra DimassiH-Index: 1
Last. Laurent KodjikianH-Index: 30
view all 8 authors...
The aim was to analyze the variations in macular vascularization on optical coherence tomography angiography (OCTA) according to the presence of intraretinal fluid (IRF) induced by exudative age-related macular degeneration (AMD). We included exudative AMD patients with IRF and/or subretinal fluid (SRF) and age-matched control eyes. All patients underwent a macular 6 × 6 mm swept-source OCTA. The mean perfusion density (MPD) and mean vascular density (MVD) were calculated in the superficial (SCP...
Source
Purpose To analyze the repeatability of vessel density (VD) measurements and manual foveal avascular zone (FAZ) measurements using optical coherence tomography angiography (OCTA) in patients with retinal vein occlusion (RVO) without macular edema. Methods The study population consisted of patients with RVO and central macular thickness l300 μm. For each subject, measurements were performed twice with a 5-minute interval. The intraclass correlation coefficient (ICC) and coefficient of variation (...
Source
#1Huijuan Wang (UM: University of Miami)
#2Huiling Hu (UM: University of Miami)H-Index: 2
Last. Jianhua Wang (UM: University of Miami)H-Index: 48
view all 6 authors...
The goal of the study was to determine the effect of different software versions on the measurement of retinal vessel densities using optical coherence tomography angiography (OCTA) in normal subje...
Source
Last. Tariq Aslam (Central Manchester University Hospitals NHS Foundation Trust)H-Index: 22
view all 5 authors...
The aim of this paper is to investigate the clinical utility of the application of deep learning denoise algorithms on standard wide-field Optical Coherence Tomography Angiography (OCT-A) images. This was a retrospective case-series assessing forty-nine 10 × 10 mm OCT-A1 macula scans of 49 consecutive patients attending a medical retina clinic over a 6-month period. Thirty-seven patients had pathology; 13 had none. Retinal vascular layers were categorised into superficial or deep capillary plexu...
Source
#1Ki-Young Kim (Kyung Hee University)H-Index: 18
#2Jong In You (Kyung Hee University)
Last. Seung-Young Yu (Kyung Hee University)H-Index: 13
view all 6 authors...
Purpose To investigate the diagnostic utility of microvascular parameters for grading the severity of diabetic retinopathy (DR) with a range of views using wide-field swept-source optical coherence tomography angiography (SS-OCTA). Methods This retrospective study grouped 235 eyes with diabetes into the five grades: diabetes without retinopathy (no-DR), mild non-proliferative DR (NPDR), moderate NPDR, severe NPDR, and proliferative DR (PDR). Foveal avascular zone (FAZ) metrics, vessel density (V...
1 CitationsSource
#1Inês Laíns (MEE: Massachusetts Eye and Ear Infirmary)H-Index: 16
#2Jay C. Wang (MEE: Massachusetts Eye and Ear Infirmary)H-Index: 9
Last. John B. Miller (Harvard University)H-Index: 26
view all 9 authors...
Abstract The advent of optical coherence tomography (OCT) revolutionized both clinical assessment and research of vitreoretinal conditions. Since then, extraordinary advances have been made in this imaging technology, including the relatively recent development of swept-source OCT (SS-OCT). SS-OCT enables a fast scan rate and utilizes a tunable swept laser, thus enabling the incorporation of longer wavelengths than conventional spectral-domain devices. These features enable imaging of larger are...
3 CitationsSource
#1Amir H. Kashani (SC: University of Southern California)H-Index: 26
#2Samuel Asanad (UMB: University of Maryland, Baltimore)H-Index: 7
Last. John M. Ringman (SC: University of Southern California)H-Index: 27
view all 12 authors...
Abstract Retinal imaging technology is rapidly advancing and can provide ever-increasing amounts of information about the structure, function and molecular composition of retinal tissue in humans in vivo. Most importantly, this information can be obtained rapidly, non-invasively and in many cases using Food and Drug Administration-approved devices that are commercially available. Technologies such as optical coherence tomography have dramatically changed our understanding of retinal disease and ...
1 CitationsSource
#1Jimmy Hong (SGH: Singapore General Hospital)H-Index: 3
#1Jimmy Hong (SGH: Singapore General Hospital)
Last. Jacqueline ChuaH-Index: 17
view all 9 authors...
We assessed the inter-visit repeatability of 15 × 9-mm2 swept-source OCTA (SS-OCTA; PLEX Elite 9000, Carl Zeiss Meditec) metrics in 14 healthy participants. We analysed the perfusion density (PD) of large vessels, superficial capillary plexus (SCP), and deep capillary plexus (DCP) as well as choriocapillaris flow voids in 2 different regions: the macular region and peripheral region. Also, retinal plexus metrics were processed further using different filters (Hessian, Gabor and Bayesian) while c...
Source