IPEM Topical Report: an evidence and risk assessment based analysis of the efficacy of tube and generator quality assurance tests on general x-ray units

Volume: 63, Issue: 24, Pages: 245011 - 245011
Published: Dec 12, 2018
Abstract
This work aims to assess the efficacy of current x-ray quality assurance (QA) testing regimes on tube and generator systems for general radiographic usage in the UK. 1393 sets of QA results data from nine UK medical physics departments were collected and analysed. Test failure rates ranged from 0% to 39% and were used to assess the likelihood of the test finding a fault. The magnitude of the recorded faults were used to assess the severity of...
Paper Details
Title
IPEM Topical Report: an evidence and risk assessment based analysis of the efficacy of tube and generator quality assurance tests on general x-ray units
Published Date
Dec 12, 2018
Volume
63
Issue
24
Pages
245011 - 245011
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.