Detectors assessment for stereotactic radiosurgery with cones

Volume: 19, Issue: 6, Pages: 88 - 98
Published: Sep 14, 2018
Abstract
The purpose of this work is to assess eight detectors performance for output factor (OF), percent depth dose (PDD), and beam profiles in a 6-MV Clinac stereotactic radiosurgery mode for cone irradiation using Monte Carlo simulation as reference. Cones with diameters comprised between 30 and 4 mm have been studied. The evaluated detectors were ionization chambers: pinpoint and pinpoint 3D, diodes: SRS, P and E, Edge, MicroDiamond and EBT3...