Measuring arterial wall perfusion using photon-counting computed tomography (CT): improving CT number accuracy of artery wall using image deconvolution.

Published on Oct 1, 2017in Journal of medical imaging
· DOI :10.1117/1.JMI.4.4.044006
Kishore Rajendran11
Estimated H-index: 11
(Mayo Clinic),
Shuai Leng53
Estimated H-index: 53
(Mayo Clinic)
+ 5 AuthorsCynthia H. McCollough76
Estimated H-index: 76
(Mayo Clinic)
Sources
Abstract
Changes in arterial wall perfusion mark the onset of atherosclerosis. A characteristic change is the increased spatial density of vasa vasorum (VV), the microvessels in the arterial walls. Measuring this increased VV (IVV) density using contrast-enhanced computed tomography (CT) has had limited success due to blooming effects from contrast media. If the system point-spread function (PSF) is known, then the blooming effect can be modeled as a convolution between the true signal and the PSF. We report the application of image deconvolution to improve the CT number accuracy in the arterial wall of a phantom and in a porcine model of IVV density, both scanned using a whole-body research photon-counting CT scanner. A 3D-printed carotid phantom filled with three concentrations of iodinated contrast material was scanned to assess blooming and its effect on wall CT number accuracy. The results showed a reduction in blooming effects following image deconvolution, and, consequently, a better delineation between lumen and wall was achieved. Results from the animal experiment showed improved CT number difference between the carotid with IVV density and the normal carotid artery after deconvolution, enabling the detection of VV proliferation, which may serve as an early indicator of atherosclerosis.
📖 Papers frequently viewed together
49 Citations
1 Citations
46 Citations
References20
Newest
#1Willi A. Kalender (FAU: University of Erlangen-Nuremberg)H-Index: 100
#2Daniel Kolditz (FAU: University of Erlangen-Nuremberg)H-Index: 13
Last. Evelyn WenkelH-Index: 17
view all 7 authors...
X-ray computed tomography (CT) has been proposed and evaluated multiple times as a potentially alternative method for breast imaging. All efforts shown so far have been criticized and partly disapproved because of their limited spatial resolution and higher patient dose when compared to mammography. Our concept for a dedicated breast CT (BCT) scanner therefore aimed at novel apparatus and detector design to provide high spatial resolution of about 100 μm and average glandular dose (AGD) levels o...
52 CitationsSource
#1Steven M. Jorgensen (Mayo Clinic)H-Index: 15
#2Mark Korinek (Mayo Clinic)H-Index: 2
Last. Erik L. Ritman (Mayo Clinic)H-Index: 77
view all 8 authors...
Early atherosclerosis changes perfusion of the arterial wall due to localized proliferation of the vasa vasorum. When contrast agent passes through the artery, some enters the vasa vasorum and increases radiopacity of the arterial wall. Technical challenges to detecting changes in vasa vasorum density include the thin arterial wall, partial volume averaging at the arterial lumen/wall interface and calcification within the wall. We used a photon-counting spectral CT scanner to study carotid arter...
6 CitationsSource
#1Zhicong Yu (Mayo Clinic)H-Index: 11
#2Shuai Leng (Mayo Clinic)H-Index: 53
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 8 authors...
Photon counting detector (PCD)-based computed tomography (CT) is an emerging imaging technique. Compared to conventional energy integrating detector (EID)-based CT, PCD-CT is able to exclude electronic noise that may severely impair image quality at low photon counts. This work focused on comparing the noise performance at low doses between the PCD and EID subsystems of a whole-body research PCD-CT scanner, both qualitatively and quantitatively. An anthropomorphic thorax phantom was scanned, and...
30 CitationsSource
#1Shuai Leng (Mayo Clinic)H-Index: 53
#2Zhicong Yu (Mayo Clinic)H-Index: 11
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 12 authors...
An ultrahigh-resolution (UHR) data collection mode was enabled on a whole-body, research photon counting detector (PCD) computed tomography system. In this mode, 64 rows of 0.45 mm×0.45 mm detector pixels were used, which corresponded to a pixel size of 0.25 mm×0.25 mm at the isocenter. Spatial resolution and image noise were quantitatively assessed for the UHR PCD scan mode, as well as for a commercially available UHR scan mode that uses an energy-integrating detector (EID) and a set of comb fi...
46 CitationsSource
#1Zhicong Yu (Mayo Clinic)H-Index: 11
#2Shuai Leng (Mayo Clinic)H-Index: 53
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 4 authors...
: Photon-counting computed tomography (PCCT) is an emerging imaging technique that enables multi-energy imaging with only a single scan acquisition. To enable multi-energy imaging, the detected photons corresponding to the full x-ray spectrum are divided into several subgroups of bin data that correspond to narrower energy windows. Consequently, noise in each energy bin increases compared to the full-spectrum data. This work proposes an iterative reconstruction algorithm for noise suppression in...
55 CitationsSource
#1Zhicong Yu (Mayo Clinic)H-Index: 11
#2Shuai LengH-Index: 53
Last. Cynthia H. McColloughH-Index: 76
view all 7 authors...
A research photon-counting computed tomography (CT) system that consists of an energy-integrating detector (EID) and a photon-counting detector (PCD) was installed in our laboratory. The scanning fields of view of the EID and PCD at the isocenter are 500 and 275 mm, respectively. When objects are larger than the PCD scanning field of view, a data-completion scan (DCS) using the EID subsystem is needed to avoid truncation artifacts in PCD images. The goals of this work were to (1) find the impact...
37 CitationsSource
#1Ralf Gutjahr (TUM: Technische Universität München)H-Index: 7
#2Ahmed F. HalaweishH-Index: 16
Last. Cynthia H. McColloughH-Index: 76
view all 10 authors...
ObjectivesThe purpose of this work was to measure and compare the iodine contrast-to-noise ratio (CNR) between a commercial energy-integrating detector (EID) computed tomography (CT) system and a photon-counting detector (PCD) CT scanner capable of human imaging at clinical dose rates, as well as to
111 CitationsSource
#1Shuai Leng (Mayo Clinic)H-Index: 53
#2Zhicong Yu (Mayo Clinic)H-Index: 11
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 12 authors...
A high-resolution (HR) data collection mode has been introduced to a whole-body, research photon-counting-detector CT system installed in our laboratory. In this mode, 64 rows of 0.45 mm x 0.45 mm detector pixels were used, which corresponded to a pixel size of 0.25 mm x 0.25 mm at the iso-center. Spatial resolution of this HR mode was quantified by measuring the MTF from a scan of a 50 micron wire phantom. An anthropomorphic lung phantom, cadaveric swine lung, temporal bone and heart specimens ...
14 CitationsSource
#1Christoph Polster (LMU: Ludwig Maximilian University of Munich)H-Index: 3
#2Katharina Hahn (Siemens)H-Index: 4
Last. T. G. Flohr (Siemens)H-Index: 1
view all 7 authors...
Photon counting detectors in computed tomography facilitate measurements of spectral distributions of detected X-ray quanta in discrete energy bins. Along with the dependency on wavelength and atomic number of the mass attenuation coefficient, this information allows for reconstruction of CT images of different material bases. Decomposition of two materials is considered standard in today’s dual-energy techniques. With photon-counting detectors the decomposition of more than two materials become...
1 CitationsSource
#1Zhicong Yu (Mayo Clinic)H-Index: 11
#2Shuai Leng (Mayo Clinic)H-Index: 53
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 11 authors...
This study evaluated the conventional imaging performance of a research whole-body photon-counting CT system and investigated its feasibility for imaging using clinically realistic levels of x-ray photon flux. This research system was built on the platform of a 2nd generation dual-source CT system: one source coupled to an energy integrating detector (EID) and the other coupled to a photon-counting detector (PCD). Phantom studies were conducted to measure CT number accuracy and uniformity for wa...
140 CitationsSource
Cited By4
Newest
#1Mårten Sandstedt (Linköping University)H-Index: 2
#2Jeffrey F. Marsh (Mayo Clinic)H-Index: 1
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 8 authors...
OBJECTIVES To compare the accuracy of coronary calcium quantification of cadaveric specimens imaged from a photon-counting detector (PCD)-CT and an energy-integrating detector (EID)-CT. METHODS Excised coronary specimens were scanned on a PCD-CT scanner, using both the PCD and EID subsystems. The scanning and reconstruction parameters for EID-CT and PCD-CT were matched: 120 kV, 9.3-9.4 mGy CTDIvol, and a quantitative kernel (D50). PCD-CT images were also reconstructed using a sharper kernel (D60...
1 CitationsSource
#1Xugao ChenH-Index: 1
#2Jianxun ZouH-Index: 1
Last. Guowei YeH-Index: 1
view all 5 authors...
BACKGROUND The aim of this study was to explore the influence of different input arteries on the parameters of computed tomography (CT) perfusion imaging for patients with different degree of stenosis of internal carotid artery (ICA). MATERIAL AND METHODS Forty patients were enrolled in the present study and divided into mild, moderate, severe stenosis and occlusion groups respectively with each 10 patients in each group. In reconstruction of cerebral CT perfusion (CTP) images, each raw perfusio...
1 CitationsSource
#1Tania Pereira (University of Lleida)H-Index: 8
#2Àngels Betriu (Hospital Universitari Arnau de Vilanova)H-Index: 18
Last. Rui Alves (University of Lleida)H-Index: 29
view all 3 authors...
Abstract Carotid adventitia vasa vasorum neovascularization (VVn) is associated with the initial stages of arteriosclerosis and with the formation of unstable plaque. However, techniques to accurately quantify that neovascularization in a standard, fast, non-invasive, and efficient way are still lacking. The development of such techniques holds the promise of enabling wide, inexpensive, and safe screening programs that could stratify patients and help in personalized preventive cardiovascular me...
5 CitationsSource
#1Felix K. Kopp (TUM: Technische Universität München)H-Index: 12
#2Heiner Daerr (Philips)H-Index: 9
Last. Peter B. Noël (TUM: Technische Universität München)H-Index: 30
view all 14 authors...
The purpose of this study was to investigate a preclinical spectral photon-counting CT (SPCCT) prototype compared to conventional CT for pulmonary imaging. A custom-made lung phantom, including nodules of different sizes and shapes, was scanned with a preclinical SPCCT and a conventional CT in standard and high-resolution (HR-CT) mode. Volume estimation was evaluated by linear regression. Shape similarity was evaluated with the Dice similarity coefficient. Spatial resolution was investigated via...
16 CitationsSource