Relative dosimetry with an MR-linac: Response of ion chambers, diamond, and diode detectors for off-axis, depth dose, and output factor measurements

Volume: 45, Issue: 2, Pages: 884 - 897
Published: Dec 21, 2017
Abstract
Purpose The purpose of this study was to acquire beam data for an MR ‐linac, with and without a 1.5 T magnetic field, by using a variety of commercially available detectors to assess their relative response in the magnetic field. The impact of the magnetic field on the measured dose distribution was also assessed. Methods An MR ‐safe 3D scanning water phantom was used to measure output factors, depth dose curves, and off‐axis profiles for...
Paper Details
Title
Relative dosimetry with an MR-linac: Response of ion chambers, diamond, and diode detectors for off-axis, depth dose, and output factor measurements
Published Date
Dec 21, 2017
Volume
45
Issue
2
Pages
884 - 897