Estimating the absolute flux distribution for a synchrotron X-ray beam using ionization-chamber measurements with various filters

Volume: 24, Issue: 5, Pages: 939 - 953
Published: Aug 11, 2017
Abstract
It is shown that an extensive set of accurate ionization-chamber measurements with a primary polychromatic synchrotron X-ray beam transmitted through various filter combinations/thicknesses can be used to quite effectively estimate the absolute flux distribution. The basic technique is simple but the `inversion' of the raw data to extract the flux distribution is a fundamentally ill-posed problem. It is demonstrated, using data collected at the...
Paper Details
Title
Estimating the absolute flux distribution for a synchrotron X-ray beam using ionization-chamber measurements with various filters
Published Date
Aug 11, 2017
Volume
24
Issue
5
Pages
939 - 953